![Hettich ROTANTA 460 S Operating Instructions Manual Download Page 18](http://html1.mh-extra.com/html/hettich/rotanta-460-s/rotanta-460-s_operating-instructions-manual_2127799018.webp)
EN
18/35
27 Program linking (for centrifuges with program linking only).
Several centrifugation runs can be linked together with the help of the program linking function.
27.1 Linking
programs
Program linking is only possible with programs in which run-up and braking steps have been set.
Before linking, the programs must be stored in the desired sequence either by entering the program or by
calling up the program (see Chapter "Programming").
The program locations must be consecutive (e.g. program locations 10+11+12).
1. Press
the
PROG
key in order to select the parameter PROG-Nr. The input field will be displayed inversely.
2. Set the program location for the first program (XX+) by means of the control knob .
3. Press
the
RCL
key. The centrifugation data for the selected program location will be displayed.
4. Press
the
PROG
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
5. Press
the
STO
key twice. The program will be linked and the program number for the next program location
(+XX+) will be displayed.
6. Press
the
RCL
key twice. The centrifugation data for the selected program location will be displayed.
7. Press
the
STO
key twice. The program will be linked and the program number for the next program location
(+XX+) will be displayed.
8. Keep repeating steps 6 and 7 until all programs have been linked.
9. Press
the
PROG
key to exit. The program number for the last program (+XX) will be displayed.
27.2 Changing program links
•
Call up the desired program (see Chapter "Calling up programs"), change the desired parameters (see Chapter
"Entering centrifugation parameters"), and save the modified centrifugation data to the same program location
(see Chapter "Entering / modifying programs"). As a result of saving, the program linking will be removed.
•
Re-link the programs (see Chapter "Linking programs").
27.3 Centrifugation run with program linking
•
Press the
PROG
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
•
Set the program location for the first program (XX+) by means of the control knob .
•
Press the
RCL
key. The centrifugation data for the selected program location will be displayed.
•
Press the
START
key. The rotation indicator will appear as long as the rotor is turning.
The run-up and braking steps for the program linking will be displayed:
x
PR-PART
First program (XX+)
x:
Run-up step for the first program
x
PR-PART
Next program (+XX+)
x:
Run-up step for the next program
y
x
PR-PART
Last program (+XX)
x:
Run-up step for the last program
y:
Braking step for the last program
•
When the time in the last program has expired, run-down will take place with the braking step of the last
program. If the centrifugation run is interrupted by pressing the
STOP
key, run-down will take place with the
braking step of the program running at the time.
27.4 Deleting program linking
•
Press the
PROG
key in order to display the parameter PROG-No. The input field will be displayed inversely.
•
Set the program location for the first program (XX+) by means of the control knob .
•
Press the
RCL
key. The centrifugation data for the selected program location will be displayed.
•
Press the
PROG
key twice in order to display the parameter PR-PART. The input field will be displayed inversely.
•
Press the
STO
key twice.
•
Press the
PROG
key.