EN
18/55
1
Use according to specification
The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG.
The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also
includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this
purpose. A different use or application over and above this is deemed not in accordance with the specifications. The
company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
2 Residual
risks
The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3 Technical
specifications
Manufacturer
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
Model HAEMATOKRIT
210
Type 2104
2104-01
Mains voltage (
±
10%)
208 - 240 V 1
∼
100 - 127 V 1
∼
Mains frequency
50 - 60 Hz
50 - 60 Hz
Connected load
250 VA
220 VA
Current consumption
1.35 A
2.2 A
Max. capacity
24 x 2.2 ml
Allowed density
1.2 kg/dm
3
Speed (RPM)
13000
Force (RCF)
16060
Kinetic energy
850 Nm
Obligatory inspection (BGR 261)
no
Ambient conditions (EN 61010-1)
−
Set-up site
Indoors only
−
Altitude
Up to 2000 m above sea level
−
Ambient temperature
2°C to 40°C
−
Humidity
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
−
Excess-voltage category
(IEC 60364-4-443)
ΙΙ
−
Pollution degree
2
Device protection class
Ι
Not suitable for use in explosion-endangered areas.
EMC
−
Emitted interference (suppression
of radio interference)
EN 55011,
Group 1, Class B
EN 61000-3-2
EN 61000-3-3
FCC Class B
−
Interference immunity
EN 61000-6-1
----
Noise level (dependent on rotor)
≤
58 dB(A)
Dimensions
−
Width
275 mm
−
Depth
330 mm
−
Height
247 mm
Weight
approx. 10 kg