12 

Maintenance and servicing 

 

The device can be contaminated. 

 

 

Pull the mains plug before cleaning. 

For safety reasons you must wear gloves and a respiratory mask when cleaning the equipment used for 
blood processing. 

Before any other cleaning or decontamination process other than that recommended by the manufacturer is 
applied, the user has to check with the manufacturer that the planned process does not damage the device. 

 

Centrifuges, rotors and accessories must not be cleaned in rinsing machines. 

 

They may only be cleaned by hand and disinfected with liquids.  

 

The water temperature must be between 20 

– 25°C, 68°F – 77°F. 

 

Only detergents/disinfectants may be used which: 

  have a pH between 5 - 8  

  do not contain caustic alkalis, peroxides, chlorine compounds, acids and alkaline solutions 

 

In order to prevent appearances of corrosion through cleaning agents or disinfectants, the application guide from 
the manufacturer of the cleaning agent or disinfectant are absolutely to be heeded. 

 

Certain preservation agents, in azide-free saline solutions can cause long term damage effects to the plastic parts 
in the device. Regular cleaning prevents salt deposits and lengthens the life span of these parts. 

12.1 

Centrifuge 

 

The following must be carried out daily: 

  Check the hoses and their attachments. The hoses must not be cracked or blocked and must be securely 

attached. The used saline solution must be allowed to flow off freely through the drainage pipe. 

  The centrifugal area must be clean and free from dried up salt crystals and other deposits. Clean the 

centrifugation area, the drain gutter and the covering ring with a moist cloth or sponge. The drain gutter and 
the covering ring can be removed from the centrifugation area for cleaning, refer to Chapter "Removal of the 
drain gutter". 

 

Check the filling volume of the saline solution (see the “Adjusting the filling volume” chapter). 

  The system must be rinsed through with distilled water to avoid formation of salt crystals, refer to Chapter 

"System with rinsing cleaning program (Clean Program)". 

 

It is important that the hoses are kept clean and free from dried up salt crystals and other deposits. 

 

The system must be regularly cleaned, refer to Chapter "Clean system with cleaning solution". It is recommended 
to clean the appliance at least once a week. 

 

Clean the centrifuge housing and the centrifuging chamber regularly, using soap or a mild detergent and a damp 
cloth if required. This serves as hygiene protection and prevents corrosion caused by impurities. 

 

Ingredients of suitable detergents: 
soap, anionic tensides, non-ionic tensides. 

 

After using detergents, remove the detergent residue by wiping with a damp cloth.  

 

The surfaces must be dried immediately after cleaning.  

 

Lightly  rub  the  rubber  seal  of  the  centrifuge  chamber  with  talcum  powder  or  a  rubber  care  product  after  each 
cleaning. 

 

Surface disinfection: 

  If infectious materials penetrates into the centrifugal chamber this is to be disinfected immediately. 

  Ingredients of suitable disinfectants: 

ethanol, n-propanol, isopropyl alcohol, glutardialdehyde, quaternary ammonium compounds. 

  After using disinfectants, remove the disinfectant residue by wiping with a damp cloth. 

  The surfaces must be dried immediately after disinfecting. 

 

Removal of radioactive contaminants: 

  The agent must be specifically labelled as being an agent for removing radioactive contaminants. 

  Ingredients of suitable agents for removing radioactive contaminants: 

anionic tensides, non-ionic tensides, polyhydrated ethanol. 

  After removing the radioactive contaminants, remove the agent residue by wiping with a damp cloth. 

  The surfaces must be dried directly after removing the radioactive contaminants. 

 

The centrifuging chamber is to be checked for damage once a year.  

 

If damage is found which is relevant to safety, the centrifuge may no longer be put into operation. In this case, 
notify Customer Service.  

 

Summary of Contents for 1008-00

Page 1: ...2017 by Hettich AG page 1 CW II OM rev1 0 Waschzentrifuge Helmer UltraCW II Waschzentrifuge Hettich Rotolavit II Operating Instructions...

Page 2: ...Centrifugeuse lave cellule Helmer UCW II 0000020 num ro de serie Dichiariamo sotto la nostra unica responsabilit che il dispositivo medico diagnostico in vitro Centrifuge per lavaggio Hettich Rotolav...

Page 3: ...d valuation du risique Procedimento di valutazione del rischio EN ISO 14971 Konformit tsbewertungsverfahren Conformity assessment procedure Proc dure d valuation da la conformit Procedimento di valut...

Page 4: ...AG Seestrasse 204a CH 8806 Baech Switzerland phone 41 0 44 786 80 20 info hettich ch www hettich ch 2017 by Hettich AG All rights reserved No part of this publication may be reproduced without the pri...

Page 5: ...e of malfunctions and irregularities 11 4 1 Remaining risks 11 4 2 Switching off the device in an emergency 11 4 3 Emergency release 11 5 Technical specifications 12 6 Unpacking the centrifuge 13 6 1...

Page 6: ...ss 25 9 4 7 AGIT process 25 9 4 8 LOOP process 26 9 4 9 CHECK process 26 9 5 Add a new program 27 10 Adjusting 29 10 1 Entering the rotor type 29 10 2 Calibrating the filling volume 29 10 3 Audible si...

Page 7: ...2017 by Hettich AG page 7 CW II OM rev1 0 Hettich AG Laborapparate phone 41 44 786 80 20 Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch 16 1 Rotors and accessories 37...

Page 8: ...al damage and injury to personal Symbol in this document This symbol refers to important circumstances Symbol on the device and in this document Beware of biohazard Symbol on the device and in this do...

Page 9: ...less condition of the device and for its proper operation in accordance to the specifications is responsible for ensuring that persons who are to operate or service the device are qualified to do this...

Page 10: ...and the legal regulations on accident prevention you should also follow the recognised professional regulations for working in a safe and professional manner These operating instructions should be re...

Page 11: ...ions or damage immediately take the device out of work and inform your superior You can find on correcting malfunctions in chapter 13 4 1 Remaining risks The device is built according to the state of...

Page 12: ...yes no yes optional car battery DC input no no 11V 30VDC width mm inch 330mm 13inch depth mm inch 480mm 18 9inch height mm closed 280mm 11inch height mm open 580mm 22 9inch weight kg lb 24 4 53 8 24...

Page 13: ...nce for damage If you notice deviations from the delivery note damage or irregularities do not put the appliance into operation but inform the haulage company and your supplier If it is possible keep...

Page 14: ...7 1 Connections 1 mounting bracket for the Power supply 2 ON OFF main switch 3 Ethernet interface 4 DC power input 1 A 5 fuse fuse holder 6 inlet 1 Saline 7 inlet 2 solution 2 1 8 outlet Drain A solut...

Page 15: ...on fig 7 1 pos 6 and dip the other side with the metal part into the container with the saline solution If the fill hose is to short for your installation and it is not possible to move closer the ca...

Page 16: ...creen occurs push the lid open button and open the lid chapter 8 2 pos 5 remove the transportation safety device from the top of the rotor and keep it on a save place 8 Operation settings 8 1 Overview...

Page 17: ...process of the program x 2 edit process x 3 back without saving to select a program x 4 add process x 5 navigate through the existing process of the program x 6 rename the program x 8 5 System Setting...

Page 18: ...em settings 2 last or next month the day with a yellow background has some saved datas runs 8 5 2 Time settings Description Login wheel adjuster for date and time 1 accept changes 2 discard changes an...

Page 19: ...lid x 2 fill in solution refill undefined x 3 start calibration x 4 back to service menu without saving x 5 select pump x 6 select target for measurement x 7 indicate filling level x 8 target measure...

Page 20: ...w hettich ch 8 6 3 Device settings Description Login 1 save changes x 2 back to service menu without saving x 3 change device name x 4 activate deactivate pumps x 5 adjust motor speed limit x 6 add ne...

Page 21: ...utton Select a program To add a new program see chapter 0 Each program must be validated by the user with the custom tube If the tube will be changed then the program must be revalidated After select...

Page 22: ...pin 30sec 3500 susp 3ml spin 30sec wash and anti 9 3 1 flush 1 This preinstalled system program is especially to flush the external and also the internal tubing system for that the first FILL process...

Page 23: ...3500RPM DECANT 300RPM AGIT 19x LOOP 2x 9 3 4 agit and spin The process values are AGIT 19x SPIN 30sec 3500RPM 9 3 5 decant The process values are DECANT 300RPM 9 3 6 spin 30sec 3500 The process value...

Page 24: ...ological saline solution is injected directly into the tubes in order to receive a good resuspension of the cells The best results are with a speed of 900RPM for both kind of rotor types The device wi...

Page 25: ...number of revolutions per minute is adjustable from 0RPM to 2500RPM The excess is decanted at the selectable rotation speed During decanting the rotor turns in the opposite direction of the normal ro...

Page 26: ...n add every other process except a LOOP Process that means you can program to wash out cells with two cycles unstead of tree with a DECANT process with a speed of a e 320RPM so that some small amount...

Page 27: ...t the name push at the top In this example the first process is FILL 1 push to choose the FILL 1 symbol Below with the bar you can choose the filling capacity and the rotation speed at which it should...

Page 28: ...add more process push By selecting a process you can switch the process to the previous process position To switch to left push Or after the next process position to right To edit a process select it...

Page 29: ...underneath the injection tube inside from the lid and push the button start calibration cha 8 6 1 pos 3 Adjust the reading from the cylinder at the target measure cha 8 6 1 pos 8 with the buttons or...

Page 30: ...rying operating hours Go into the system settings start screen cha 8 2 pos 2 Go into the service menu system settings cha 8 5 pos 2 11 Centrifugation of materials or mixtures of materials with a densi...

Page 31: ...pter Removal of the drain gutter Check the filling volume of the saline solution see the Adjusting the filling volume chapter The system must be rinsed through with distilled water to avoid formation...

Page 32: ...After removing the radioactive contaminants remove agent residue by rinsing with water only outside of the centrifuge or wipe off with a damp cloth The rotors and accessories must be dried directly a...

Page 33: ...olume ml and speed from the FILL process in the used program Clean the injection nozzles Pellet does not form on the base of used tube The revolutions per minute during the centrifugation run for the...

Page 34: ...cked Open or check button was pressed but the lid could not be opened Lid mechanically blocked 31 Unlocking failed Open or check button was pressed but the lock could not be unlocked Motor was still r...

Page 35: ...incurred for cleaning and disinfection are to be charged to the customer We ask for your understanding in this matter 15 Storage Before storage the device it should be decontaminated and cleaned to p...

Page 36: ...W II OM rev1 1 Hettich AG Laborapparate phone 41 44 786 80 20 Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch Log serialnumber ______________ inventorynumber _____________...

Page 37: ...tube Kapazit t capacity ml 3 5 Ma e dimensions x L mm 10 x 75 12 x 75 Anzahl p Rotor number p rotor 12 12 Drehzahl speed RPM 3500 2 RZB RCF 1438 Radius radius mm 105 CW 1024 A E2197 Dekantierrotor 24...

Page 38: ...ml 3 5 Ma e dimensions x L mm 10 x 75 12 x 75 Anzahl p Rotor number p rotor 12 12 Drehzahl speed RPM 3500 2 RZB RCF 1438 Radius radius mm 105 1018 A E2197 Dekantierrotor 24 fach Decant rotor 24 times...

Page 39: ...Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch 16 1 3 Document History rev date supersession Revision description 1 0 23 May 2017 01 06 Contents desc of preloaded progra...

Reviews: