![Hettich 1008-00 Operating Instructions Manual Download Page 26](http://html1.mh-extra.com/html/hettich/1008-00/1008-00_operating-instructions-manual_2127765026.webp)
© 2017 by Hettich AG
page 26
CW II-OM-rev1.1
Hettich AG
Laborapparate
phone:
+41 44 786 80 20
Seestrasse 204a
e-mail:
CH-8806 Baech
internet:
www.hettich.ch
9.4.8
LOOP process
Loop.
This process makes only sense as a following
process from minimum one other process. The
numbers of loops are adjustable from 1x to 100x.
The default-value is 3x. After the foregone process is
finished, all foregone processes will be repeated with
the adjustable value.
If the foregone process is a spin-process with a duration-time of 2 hours and the loop-process is adjusted
to 1x, then the program will repeat the spin-process one time with the duration of 2 hours, that means that
the device spins for 4 hours.
After the LOOP Process you can add every other process except a LOOP Process, that means you can
program to wash out cells with two cycles unstead of tree with a DECANT process with a speed of a.e.
320RPM so that some small amount of solution will stay into the tubes and after the LOOP process you add
the same processes again but the speed from the DECANT process is a.e. 370RPM so that the tubes are
empty of the Solution.
9.4.9
CHECK process
Check, pause.
This process makes only sense as a following
process from minimum one other process. After the
foregone process is finished the program will be
paused and the lid will open. The user can check the
probes or can add some other fluids with a pipette to
the probes. When the lid will be closed, then the
program will continue.
If the forgone processes are a wash-cycle you can add during the check-process the anti-human globulin
serum, the following processes are if nessecary a AGIT-process and a SPIN-process