Argyll II

™ User/Service Manual 

OPERATION 

 

  

Image 1DA  
 
(NB: use hand to check ensure appropriate clearance when profiled as below. Add more pressure 
on the pump if required in order to achieve this. NB: Weight setting is a guide only and cell 
pressure must also be taken into consideration for individual patients). Clinical judgement must 
always overrule user guidance. 
 

 

Image 2DA 

 

Installation 

1.  Remove all packaging & place the pump either at the end of the bed on its hooks or on a 

flat surface. 

2.  Ensure that there are no sharp or hard objects on the bed before unfolding the mattress & 

laying out flat. The air inlet should be located at the feet side of the bed. 

3.  Use the Rapid Inflator (if available) to inflate the mattress before connecting the air tubes to 

the pump. The tubes should be connected securely in place without any signs of leakage. If 
a Rapid Inflator is not available, the mattress will take approximately 50 minutes to inflate. 

NB: 

The pump will not operate dynamically until the inflation setting is reached. 

4.  Connect the pump to the socket & switch on.  
5.  Adjust the mode & pressure by the users condition & weight to suit. 
6.  In order for the alarm to activate, the system must be switched on & powered up for at least 

5 minutes. This should give the internal battery on the PCB sufficient time to charge. 

Summary of Contents for ARG-121-2

Page 1: ...ervice Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattres...

Page 2: ...2 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual...

Page 3: ...OPERATION 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair...

Page 4: ...ght Do not use phenolic based cleaning solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags s...

Page 5: ...under patient weight as is often the case Application This product is to be used for those patients who have existing tissue damage or pressure ulcers It can also be used as a high risk preventative i...

Page 6: ...gly to appropriately accommodate see below Unless the clinical condition dictates it is not advised to adjust the bed profile beyond 45 The Argyll II will continue to operate beyond this but may cause...

Page 7: ...on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting the...

Page 8: ...Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed th...

Page 9: ...tance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 50...

Page 10: ...n trained by Herida Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This...

Page 11: ...is should be done via a pressurised system soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear...

Page 12: ...fuse is F0 5A 250V When replacing the fuse ensure a similar type is used and that the fuse is disposed of in the appropriate manner During normal working when the emergency deflation is needed switch...

Page 13: ...uding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is sou...

Page 14: ...14 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual P09 3 Way Pump...

Page 15: ...gainst electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electric...

Page 16: ...tors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the us...

Page 17: ...1HE Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of in an envir...

Reviews: