Argyll II

™ User/Service Manual 

TROUBLE SHOOTING 

 

No.

 

Issue

 

Possible Cause

 

Possible Resolution

 

 
 
 
 

 

 
 
 
 

 

No power 

  The plug is not 

properly  connected. 

  Power Cut. 

  The fuse is broken. 

  The machine / system 

is  broken. 

  Make sure the power is 

correctly connected and the 
voltage is normal. 

  Await re-connection of 

power / look at generator. 

  Check the fuse, replace if 

broken. 

  Call the local distributor for 

repair or replacement. 

 
 
 
 

 
 
 
 

Air mattress leakage / 
Alarm Sounding 

  The mattress is not 

connected to the 
pump  correctly. 

  Air tube is disconnected 

or  damaged. 

  CPR not attached properly 

  Check the tube connectors. 

  Repair the air tubes or call 

for replacement. 

  Re-attach CPR plug 

 

 

 

Mattress is too hard 

or  too soft 

  The air pressure is not 

adjusting to required 
level. 

  Adjust the pump pressure 

according to the user 
weight until suitable and 
comfortable. 

 
 
 
 
 

 

 
 
 
 

 

No full inflation 

cycle after 

initialisation 

  Air hoses are loose or 

not  connect properly. 

  CPR is loose or 

not connected. 

  Air pump has not enough 

air  pressure. 

  Check all the air hoses and 

connectors. 

  Connect the CPR tightly. 

  Adjust the pressure on 

the pump. If it is showing 
maximum pressure full 
inflation has not been 
achieved, the pump has a 
problem. Ask for repair or 
replacement. 

 

 

 

Pump has 

strong  vibration 

  The pump is not located on 

a  flat area or not hooked 
properly on the bedframe. 

 

  Place the pump on flat area 

or  hook  it  in  a  suitable 
position on the bedframe. 

 
 
 
 

Summary of Contents for ARG-121-2

Page 1: ...ervice Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattres...

Page 2: ...2 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual...

Page 3: ...OPERATION 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair...

Page 4: ...ght Do not use phenolic based cleaning solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags s...

Page 5: ...under patient weight as is often the case Application This product is to be used for those patients who have existing tissue damage or pressure ulcers It can also be used as a high risk preventative i...

Page 6: ...gly to appropriately accommodate see below Unless the clinical condition dictates it is not advised to adjust the bed profile beyond 45 The Argyll II will continue to operate beyond this but may cause...

Page 7: ...on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting the...

Page 8: ...Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed th...

Page 9: ...tance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 50...

Page 10: ...n trained by Herida Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This...

Page 11: ...is should be done via a pressurised system soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear...

Page 12: ...fuse is F0 5A 250V When replacing the fuse ensure a similar type is used and that the fuse is disposed of in the appropriate manner During normal working when the emergency deflation is needed switch...

Page 13: ...uding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is sou...

Page 14: ...14 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual P09 3 Way Pump...

Page 15: ...gainst electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electric...

Page 16: ...tors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the us...

Page 17: ...1HE Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of in an envir...

Reviews: