background image

 

 

11 | 

P a g e  

A U G U S T   2 0 1 9

 

 

Argyll II

™ User/Service Manual 

MAINTENANCE 

 

At point of service the following aspects must be maintained 

to ensure that manufacturer’s 

warranty remains intact. These can be done in a non-alphabetic way, if all areas are covered:

 

 

  Ensure the pump unit and mattress / cells are fully decontaminated prior to service and are 

cleaned in accordance with the schedule provided. 

  Electrical safety checks / Pat Test 

– To be conducted annually or time of service. Please 

make sure the kettle lead and plug fuse is compliant at 3amp only.  

  Filters must be changed at point of return / before patient handover / new client and at time 

of service. 

  Rear pump fuse must be inspected and replaced if faulty 

– 0.5A only. 

  Pump casing must be opened and vacuumed internally, if any dust or particles are evident. 

Never use any water / fluid internally to clean. 

  Pump must achieve pressure within 20% tolerances as per page 9 of this manual 

– via 

calibrated sphygomometer and cell attachments. Kits available to purchase via 
manufacturer. Ensure that a rapid inflator is used at time of service to ensure that service 
and installation time is reduced to minutes. 

  Pump low pressure function is evident at 6mmHG or lower for 15-20 minutes or more. 

  Cells are to be fully inflated and tested for leakage. This should be done via a pressurised 

system, soak test over at least 3 cycles or under weights to replicate patient in situ.  

  The tubing and connectors that run down the side of the mattress, must be clear of kinks / 

debris or must not be snapped in any way. These should be running freely with air 
distribution to each cell and have no leakage or deterioration of surface. A spring is situated 
at the head section of the mattress tubing at to avoid kink in the critical area. 

  The none return valve holding the 3 head cells must be functioning correctly. This is evident 

via 3 solid static head cells / none alternating during cycle.     

 

A functional “soak test”, ensuring that the tactile panel / pump and mattress operates fully in 
synergy. This is done by attaching and inflating the mattress to full pressure and then 
increasing / decreasing both weight and other panel / button functions. 

  The key lock button is effective by release / addition, holding 3 seconds or more to activate.  

  CPR is to be checked functionally and deflated after all soak tests are completed. 

  Mattress Cover and cells are to be inspected for puncture, tear, rip or excessive odour post 

cleaning and replaced accordingly if defective. 

  Calibration setting as required. See page 12 for further details. 

 
Storage 

Good ventilation will be required when storing the product & it should always be kept dry & away 
from direct sunlight. Temperature should be maintained in the region of 22°C ± 3°C when in use. 
Avoid repeat exposure to extreme temperatures. 

 

Summary of Contents for ARG-121-2

Page 1: ...ervice Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattres...

Page 2: ...2 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual...

Page 3: ...OPERATION 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair...

Page 4: ...ght Do not use phenolic based cleaning solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags s...

Page 5: ...under patient weight as is often the case Application This product is to be used for those patients who have existing tissue damage or pressure ulcers It can also be used as a high risk preventative i...

Page 6: ...gly to appropriately accommodate see below Unless the clinical condition dictates it is not advised to adjust the bed profile beyond 45 The Argyll II will continue to operate beyond this but may cause...

Page 7: ...on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting the...

Page 8: ...Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed th...

Page 9: ...tance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 50...

Page 10: ...n trained by Herida Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This...

Page 11: ...is should be done via a pressurised system soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear...

Page 12: ...fuse is F0 5A 250V When replacing the fuse ensure a similar type is used and that the fuse is disposed of in the appropriate manner During normal working when the emergency deflation is needed switch...

Page 13: ...uding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is sou...

Page 14: ...14 P a g e A U G U S T 2 0 1 9 Argyll II User Service Manual P09 3 Way Pump...

Page 15: ...gainst electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electric...

Page 16: ...tors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the us...

Page 17: ...1HE Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of in an envir...

Reviews: