Page 20 of 21 Digilance V
26 November 2018
File name:
Digilance V_1.01
Version: 1.01
9 Technical data
Measurement Application
Single or Multi-dip temperature measurement
Measurement Channel
One analog input
Sample Rate
10 samples/second
Measurement Range
Type S (Pt 10%Rh / Pt)
Type R (Pt 13%Rh / Pt)
Type B (Pt 30%Rh / Pt 6%Rh)
Type K (Ni-Cr / Ni)
Type D (W / Re)*
400 to 1768°C (752 to 3214°F)
400 to 1768°C (752 to 3214°F)
400 to 1820°C (752 to 3308°F)
400 to 1372°C (752 to 2502°F)
400 to 2300°C (752 to 4172°F)
According to
IEC 60584
*ASTM 988
Thermocouple Tables
IPTS 48 / IPTS 68 / ITS 90
Thermocouple Accuracy
Better than +/- 1°C between 0°C and 50°C ambient temperature
Display Resolution
Configurable 1° or 0.1°
Display Units
°C or °F
Data Storage
500 measurement results with heat number and date/time storage.
Transfer of stored data using wireless technology.
Power Supply
Lithium Ion Battery Pack (3.5Ah – 7.2V)
Charging Voltage
12V DC, 2A MAX
Battery Life
40 operating hours (nominal)
Wireless Communication
(optional)
Range of 100 meters - line of sight
Frequency hopping 2.4 GHx ISM unlicensed band technology
Max 10 mW transmit power
Operating Ambient
Temperature Range
-20°C (-4°F) to 60°C (140°F), max. 90% RH non-condensing
Environmental Protection
IP55 (with lance installed)
Size
Weight (without lance)
1.9kg (4.2lbs)