Page 10 of 21 Digilance V
26 November 2018
File name:
Digilance V_1.01
Version: 1.01
4 Standard Operation
The four keys on the Digilance front panel are used to interact with the device:
Key Primary Functions
•
Enter
•
Power On/Off
•
Cancel
•
Exit
•
Access Main Menu
•
Scroll Up
•
Scroll Down
4.1 First Use
•
To Activate the Digilance for the First Time, Press and Hold
o
After the initial activation, the instrument will Power-On by movement
•
Press
to the access the Main Menu and the following options
o
History
o
Configuration Login
o
Instrument Info
o
Battery Info
o
Wireless Info (optional)
•
Prior to using the instrument, it is recommended to confirm the following instrument settings (See
Section ‘
’):
Configuration Section ID
Function
Thermocouple Type
1.1
S / B / K / R / D
Thermocouple Table
1.2
ITS90 / IPTS68 / IPTS48
Temperature Units
1.3
°C / °F
Time
1.4
Time Value Set in 24hr format
Date
1.5
Date Value Set in DD/MM/YY format
Measurement Mode
2.3
The Measurement Cycle is Reset dependent on the
Measurement Mode
•
DIP: Reset after the Thermocouple is replaced – For
use with Single-Use Probes
•
MULTI: Reset when the Temperature drops by the
amount of the Stop Difference below the Start Condition
– For use with Multi-Use Probes