background image

 

 Support Chinese or English operation menu.  

  With  the  function  of  temperature  compensation,  water-proof  and 

dust-proof, equiped with the dust filter. 

 With the temperature and humidity detection. 

3.Technical parameter

 

Product

 

Multi gas and dust particle detector 

Measuring 

range

 

Reffering to the gas list 

Resolution

 

Reffering to the gas list 

Gas type

 

At most for 5 sensors, the gas type will be according to your requirements. 

Temperature 

and 

humidity

 

Temperature

-40 

  120

℃,

Humidity

0-100%RH 

Testing 

mode

 

Pump-sunction

inner pump, adjustable flow rage(500-1000mL L/min), workable for 

the vacuum environment. 

Accuracy

 

≤±3%F.S 

Linearity

 

≤±1%F.S 

Response 

time

 

≤20s

T90

 

Zero drift

 

≤±1%

F.S/year

 

Recovery 

time

 

≤20s 

Repeatability

 

≤±1%F.S 

Testing 

mode

 

Realtime detecting or Timing detecting is adjustable. 

Data 

Storage

 

Automatic storage or manual storage is adjustable; About 100000 group of data 

capacity. 

Summary of Contents for OC-1000

Page 1: ...OC 1000 Gas and dust particle detector Operating Manual Henan Oceanus Import Export Co Ltd...

Page 2: ...1 Catalogue Notice 2 Description 4 Feature 4 Parameter 5 Structure 7 Operation instruction 7 Operation interface 9 Remarks 12 Fault Treatment 13 Data logger funct ion 14...

Page 3: ...ss 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on Treatment for the over measuring range Please av...

Page 4: ...g the free calibration service once in a year About the Backlight instruction OC 1000 with the electricity saver procedure when the device without any operation within 30s the detector will turn on th...

Page 5: ...PS industrial grade color screen displaying each parameter It is able to select gas concentration unit of PPM VOL mg m3 For 1 5 gas detecting supporting PM2 5 dust particle sensor and temperature sens...

Page 6: ...Gas type At most for 5 sensors the gas type will be according to your requirements Temperature and humidity Temperature 40 120 Humidity 0 100 RH Testing mode Pump sunction inner pump adjustable flow...

Page 7: ...rget Gas List Some gas not list if need pls contact manufacturer Gas Range Accuracy Resolution Reponse time Ex 0 100 LEL 3 F S 0 1 LEL 20s CH4 0 100 LEL 3 F S 0 1 LEL 20s O2 0 30 VOL 3 F S 0 01 VOL 20...

Page 8: ...7 4 Structure 5 Operation instruction 5 1 Keys Total seven keys Back OK run stop and print Three interfaces Testing interface Menu interface Setting interface...

Page 9: ...off Press 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on 5 2 Power on When power off long press ke...

Page 10: ...nt interface as the following picture displaying one gas two gas three gas four gas orderly As the above displaying time temperature pump running battery and humidity Take A for example for O3 CO CH2O...

Page 11: ...an turn on or off some one or more channel gas With the Pump flow rate can adjust the pump flow rate only need to setting under vaccum enviroment default is 30 With Temperature correction and Humidity...

Page 12: ...re enter into the target gas calibration interface slowly turn on the calibrate gas and control the flow rate within 500ml min observing the detector gas concentration now should be rising when the va...

Page 13: ...n is this period can set the timing cycle and round cycle time between the twice timing detecting Setting the timing detection need to back to the Detecting interface to press the Run stop key turn on...

Page 14: ...se to normal value If still not normal then zero calibrate in clean air Response weak Reason Lower level of oxygen 5 VOL High pressure pump not work The sensor useful life end or other faults Solution...

Page 15: ...if not then need to change the charger if is 5V then need to send back to manufacturer for testing 9 Data logger OC 1000 with the data logger function the charging port is the data output port the ope...

Page 16: ...be in this period range 6 Click the search can find the above period gas concentration for the according channel gas the data also could be saved as the EXCEL form by click the Read Data Remark Data...

Reviews: