92
QC lockout determined by number of
measurements
Measurements (2 - 200)
1 0 0
Lockout
Interval
1
5
2
8
7
4
6
9
3
0
11
55
22
88
77
44
66
99
33
00
1
2
3
FIGURE 3-74
If Measurements was selected in the QC
lockout menu (see
for the QC lockout based on the number of
patient or STAT test measurements
performed is displayed.
The numbers in brackets show the lowest
and the highest number of tests that can be
set.
a) Enter the number of measurements after
which QC lockout will occur (1) via the
Digit buttons (2). Press the Confirm
button (3).
b) If the reminder value needs to be
Otherwise there will be a return to
1 No. of measurements after which QC lockout
occurs
2 Digit button
3 Confirm button
Lockout
Interval
No
No
Adjust
Adjust
Note!
New Lockout Value
Requires Adjusted
Reminder Value
Adjust?
FIGURE 3-75
c) To adjust the number of measurements
for the reminder so that it corresponds
with the new value for the number of
measurements before QC lockout, press
Adjust.
After adjusting to an acceptable value
(see
), there will be a return
to the QC Test Settings menu,
.
Otherwise, to return to
press No. The selected time/Number of
measurements until the next QC lockout
will then need to be adjusted, in order to
correspond with the defined reminder.