1-360008/G
26
It is very noisy when the
basket assembly raises
and lowers and/or when
agitating
- Faulty lift-out motor
- Verify that the lift-out motor is
operating properly
- Replace as needed
- Faulty roller bearing block
- Verify that the rail moves freely up
and down in the roller bearing block
- Verify that there is no debris in the
roller bearing block or on the rail
- Faulty agitation motor
- Verify that the agitation motor is
operating properly
- Replace as needed
- Basket assembly
- Verify that the finger screw on the
top of the basket assembly is tight
- Basket assembly bearing
-
Verify that the plastic roller bearings
at the bottom of the basket assembly
are operating properly
-
Replace as needed
Thawing Cycle
Thawing cycle will not
start
- Faulty push button
- Verify Start/Stop button is activating
the timing cycle; replace as needed
- Faulty circuit board
- Verify wiring connections from
circuit board to motor assemblies
- Verify that the circuit board is
securely mounted to the chassis
studs
- Replace circuit board as needed
- Faulty positioning switches
- Verify that the two positioning
micro- switches at the roller wheel
are operating properly