D. Accessing the Chamber Set Temperature
To alter the chamber temperature setting:
1) Press and hold the SET key to display the current set temperature input value.
2) Press either the Up or Down keys to alter the set point as desired. If the set point
temperature will not change, the Set Point Lockout is in the lockout mode - refer to
"Accessing the Set Point Lockout" to unlock this program value.
3) Release the SET key to complete the chamber temperature setting change.
E. Accessing the Low and High Alarm Programs.
To alter the Low and/or High Alarm settings:
1) Press and hold the Up and Down arrow keys until a controller program is displayed. If the
"CnFg” prompt appears, the Controller Program Lockout will need to be unlocked - refer to
"Accessing the Controller Program Lockout" to unlock this program value.
2) Press the Up or Down arrow key until the “ALO” program appears for the Low Alarm, or
“Ahl” for the High Alarm program.
3) While holding the SET key down, press the Up or Down arrow keys to alter the set point
value as desired.
4) To exit back out to the main operating mode, release the SET key and press both the UP and
Down arrows until the operating temperature appears on the digital display.
F. Chamber Drain System