Helios PD2-6400 High Voltage & High Current Meter
Instruction Manual
39
Advanced Features Menu & Display Messages
Display
Parameter
Action/Setting
C Hi
Current high
Calibrate high current input
(e.g. 5 A)
Error
Error
Error, calibration not
successful, check signal or
programmed value
Ch-V
Channel V
Perform calibration on
channel V
V CAL
Voltage
calibration
Calibrate voltage input
V Lo
Voltage low
Calibrate low voltage input
(e.g. 0 V)
V Hi
Voltage high
Calibrate high voltage input
(e.g. 100 V)
Error
Error
Error, calibration not
successful, check signal or
programmed value
Diag
Diagnostics
Display parameter settings
LED t
LED test
Test all LEDs
Info
Information
Display software number
and version
erase
Erase
Erase MeterView Pro
software stored in meter’s
memory
Noise Filter (
filter
)
The noise filter is available for unusually noisy signals
that cause an unstable process variable display. The
noise filter averages the input signal over a certain
period. The filter level determines the length of time
over which the signal is averaged. The filter level can
be set between 2 and 199. The higher the filter level,
the longer the averaging time and so the longer it
takes the display to settle to its final value. Setting the
filter level to zero disables the filter function.
Noise Filter Bypass (
bypass
)
The noise filter bypass changes the behavior of the
meter so that small variations in the signal are filtered
out but large abrupt changes in the input signal are
displayed immediately. The bypass value determines
the minimum amount of signal change to be displayed
immediately. All signal changes smaller than the
bypass value are filtered or averaged by the meter.
The noise filter bypass may be set between 0.1 and
99.9% of full scale.
Rounding Feature (
round
)
The rounding feature is used to give the user a
steadier display with fluctuating signals. Rounding is
used in addition to the filter function.
Rounding causes the display to round to the nearest
value according to the rounding selected. See
examples below:
Rounding
Selection
Actual
Value
Display
Value
Actual
Value
Display
Value
1
12.022
12.022
12.023
12.023
5
12.022
12.020
12.023
12.025
10
12.024
12.020
12.025
12.030