E9Screen Software
EPC9 Manual
41
Mode
: Sets the
Recording Mode
(in the above example it is
Voltage Clamp
). For details
on the various recording modes see
Chapter
6. Operating Modes
on page 54
.
• Test
- Sets the
Test
mode.
• VC
- Sets the
Voltage Clamp
mode.
• CC
- Sets the
Current Clamp
mode.
When switching from voltage-clamp to current-clamp mode,
I-hold
will be set to
inject the holding current required to keep the membrane voltage constant (“gentle
switch”, see description of the Gentle CC-Switch control below, subchapter Hidden
Controls). Thus, V-mon will be identical to the original
V-membrane
commanded
under voltage clamp. In analogy, upon returning to voltage clamp,
V-membrane
will
be set according to the membrane potential under current clamp and rounded to the
next 2 mV.
Quite often, the membrane potential under current clamp will
have changed, and when returning to voltage clamp,
V-
membrane
will possibly differ from that set before switching to current-clamp mode.
Last V-membrane
can then be used to conveniently restore the original
V-membrane
value. This button is normally hidden behind the oscilloscope window; it can be seen
when closing the oscilloscope. For details see
Chapter
6. Operating Modes
on page 54
.
Starting with the version “C” of the EPC9, the current-clamp
circuitry has two speed settings (which can be toggled by a
button that is normally hidden behind the oscilloscope window; it can be seen when
closing the oscilloscope). For details see
Chapter
6. Operating Modes
on page 54
.
Filter 1
: Controls an analog 3-pole filter in
the current monitor pathway. The pop-up
menu provides for the following settings:
• Bessel 100 kHz
• Bessel 30 kHz
• Bessel 10 kHz
• HQ 30 kHz
Under most conditions a 10 kHz bandwidth is more than ample, and the filtering
reduces the high-frequency noise substantially. The HQ 30 kHz setting is selected
automatically when fast R
s
-compensation is in use; it is of little use otherwise.
Filter 2
: Controls an analog 4-pole filter for
the current monitor 2 (
I-mon 2
). Filter 2 is in
series with Filter 1. Dragging the mouse or typing allows fine adjustment from 0.1-16