7.2.9.1
Status output active state
The status output can be compared to an electrical relay that can function as make or break contact. For safety-
relevant applications, the operator will choose the break contact setting so that a power failure or failure of the
electronics can be detected like an alarm. In standard applications, the output is used as make contact.
The Status output state active state function allows the operator to define the behaviour of the status output.
Status output active
[closed]
The current active state will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:
closed
open
Pressing the “
” key confirms the entry.
7.2.9.2
Status output 1 assignment
This function allows the operator to define to which event the status output is to be assigned. The most general
assignment is the alarm assignment because all set limiting values and the self-test function are then
monitored via the status output. After selecting the Status output assignment function, press “
” to display the
current assignment.
Output 1 assigned to
[Alarm]
The current assignment will be displayed and can be changed by toggling the arrow keys.
The operator can choose between the following settings:
Flow direction recognition
o
Forward flow
o
Reverse flow
Limiting values:
o
MIN QM
o
MAX QM
o
MIN density
o
MAX density
All limiting values and error detection
o
Alarm
Pulse output 2 for custody transfer operations
o
IMP2 90°,
Pressing the “
” key confirms the entry.
When selecting the IMP2 90° setting, a second pulse output will be realised via the status output that can be
used for custody transfer operations.