12/64
med 0413 2017-12-07
Technical data
*
smallest light field and at a 30 cm working distance, depending on t
General instructions
Condensation can form when moving from a cold to a warm environment.
Wait until the unit has warmed up and is absolutely dry before using it.
Proper functioning is only guaranteed if original spare parts and
accessories from HEINE are used.
Electromagnetic Compatibility
Medical electric devices are subject to special precautionary measures with
regard to electromagnetic compatibility (EMC). Portable and mobile high
frequency communication equipment can affect medical electric devices.
This device is intended for use by medical professionals in the electro-
magnetic environment specified below. The user of this device should
assure that it is used in such an environment
The use of accessories, converters or cables other than the ones
specified by HEINE might lead to increased emission reduced electrical
immunity of the medical equipment.
The ME device may not be stacked directly near or used directly beside
other devices. If the ME device is to be operated in a stack or with other
devices, the device should be watched to ensure it operates properly in
this location.
Mains power:
100– 240 V AC, 50/60 Hz
Power consumption:
18 VA
Fuse sets:
2 x T 2A H 250 V
Life time of the lamp:
approx. 30,000 hours
Expected service life:
≥ 5 years
Minimum distance from light source to
patient:
20 cm
Working distance from light source
to patient:
30 cm
Central illuminance Ec:
≥ 45,000 lx
*
Light field diameter d
10
(at a 30 cm
working distance):
approx. 8.5 – 14 cm
Overall irradiance:
Typically 115 W/m
2
Colour temperature:
≥ 3000 K
Colour rendering index Ra:
≥ 85
Special colour index R9:
Typically 58
Environmental conditions for operation:
Temperature:
Relative humidity:
Air pressure:
+10 °C to +35 °C
10 % to 75 %
700 hPa to 1060 hPa
Environmental conditions for storage
and transport:
Temperature:
Relative humidity:
Air pressure:
-20 °C to +50 °C
10 % to 95 %
500 hPa to 1060 hPa
Protection class:
I
Operation altitude:
3000 m
Protection provided by enclosures
(IP code):
IP20
Risk group (according to DIN EN 62471) 1 (low risk)
Weight incl. wall mount:
approx. 2.6 kg
Weight of clamp mount:
approx. 0.2 kg
Length of flexible arm:
97 cm
Guidance and manufacturer’s declaration – electromagnetic emissions
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such environment.
Emission test
Compliance
Electromagnetic environment – Guidelines
RF emissions CISPR11
Group 1
EUT uses RF energy only for its internal function. Therefore, RF-emission is very low and it is
unlikely that any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The EUT is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Warning: This device is intended only for use by medical professionals. This is a device of class A
CISPR 11 in the domestic environment, this device may cause radio interference, so that it may be
necessary in this case, to take appropriate remedial measures, as e.g. orientation, new arrange-
ment or shielding of the MEG or restrict the connection to the site.
Harmonic Emissions
IEC 61000-3-2
Class A
Symmetrical three-phase EUTs and other EUTs.
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Passed