SSelect input high value
Setting range depends on measuring range [V / mA]
SSelect high display
vvalue for I
I
NP.HI.
Setting range -199999 … +999999 and decimal point
SSelect input filter:
:
(number of measuring cycles for floating average calculation)
Setting range 1 … 99 [cycles], with setting 1 the filter is turned off
SSelect start delay
aafter power on
(limit monitoring is processed only after the programmed time has expired)
Setting range 0.0 ... 99.9 [sec]
SSelect
llocal mains frequency
Mains frequency 50 Hz
Mains frequency 60 Hz
x
TThe setting range depends on the selected measuring range:
0 … 10 V
=
-0.500 … 10.500 [V]
2 … 10 V
=
1.500 … 10.500 [V]
-10 … +10 V =
-10.500 ... +10.500 [V]
0 … 20 mA
=
-0.500 … 21.000 [mA]
4 … 20 mA
=
3.000 … 21.000 [mA]
x
The functions LO.LIM and HI.LIM will limit the editable range correspondingly.
x
The setting of HHI.LIM has to be greater than the setting of LLO.LIM aat any time.
7.3 User Linearization
Menu
u
U
U
ser-L
L
inearization
Select User Linearization
Linearization OFF
Linearization ONF
Select number of interpolat
t
ion points
Setting range 3 … 12 points
Select input low value
Setting range depends on measuring range [V / mA]
Select display low value for
INP.LO.
Setting range -199999 … +999999 and decimal point.