Troubleshootimg
26
Troubleshooting
Possible errors and how to resolve
General
Malfunctions
Possible reason
Troubleshooting
Power LED does not
enlighten
No power
▪
Check size and shape of
plug and compatibility to
your electrical socket
▪
Check circuit breakers
Light diode defective
Contact your local Heidolph
Instruments distributor
No rotation function
No stirring bars in vessel
Insert stirring bars
No heating up function
Set temperature exceeded
Let device cool down
Sample temperature
lower than desired
Temperature controlled via
hotplate
Raise temperature of
hotplate
MR Hei-Standard
Error message
Possible reason
Troubleshooting
Operating indicator light
blinking 1 time
No heating function
▪
Hotplate temperature
>25 °C to set temperature
▪
Hotplate sensor defective
Contact your local Heidolph
Instruments distributor
Operating indicator light
blinking 2 times
No heating function
Set temperature external
Pt 1000 exceeded for more
than 25°C:
▪
Exothermic reaction
▪
Test setup problematic
Adjust test setup
▪
Temperature control Pt
1000 defective
Contact your local Heidolph
Instruments distributor
Operating indicator light
blinking 3 times
No heating function
▪
Rotation knob defective
▪
Temperature knob
defective
▪
Heating button defective
Contact your local Heidolph
Instruments distributor
Operating indicator light
blinking 4 times
▪
No heating function
▪
No rotating function
Motor defective
Contact your local Heidolph
Instruments distributor