713687 rev 9
7
9.4
Settings
9.4.1
1 Select Language
Use the up and down
arrow keys
to scroll through the four-
teen different languages (arranged alphabetically) until the
correct language appears. Press
select
to confirm.
If you accidently switch to a language that you don´t under-
stand, turn the main switch on while pressing the “cassette
in use” button. The dryer will then start up with British
English.
9.4.2
3 Process Alarm
Use the up and down
arrow keys
to set whether the alarm
is active or not. Press
select
to confirm. For alarm sound
settings, see “9.4.4 Buzzer”.
9.4.3
4 Temp. Unit
°C and °F appear on the screen. Use the up and down
arrow keys
to select the desired temperature unit. Confirm
the selected unit by pressing
select
.
9.4.4
5 Buzzer
The sound effects can be muted/activated by using the
arrow keys
to select no/yes. Press
select
to confirm.
9.4.5
7 Basic PIN Code
You can change the PIN code to prevent unauthorized ac-
cess to the basic settings, as described in table in chapter
9.4. To do this use the up and down
arrow keys
to change
the first digit, press
select
to confirm. Continue with the
second, third and fourth digit. When completed press
select
to confirm the new four digit PIN code.
9.4.6
8 Short Distance Limit
The original settings for the correct short distance limit is
55 cm. You can change the value for the short distance
limit from 35 to 95 cm. Settings higher than 80 cm are not
recommended. Press
select
to confirm.
9.4.7
9 Long Distance Limit
The orginal settings for the correct long distance limit is
65 cm. You can change the value for the long distance limit
from 40 to 130 cm. Settings higher than 100 cm are not
recommended. Press
select
to confirm.
9.4.8
15 Temperature Ctrl
You can choose if you want to turn the temperature control
(pyrometer) on or off permanently.
Note!
If you turn the temperature control off in settings
mode you cannot activate it in program mode.
Temperature graphs and values will not be accessible when
this feature is turned off. Instead the software will work with
power levels, that is percentage of maximum capacity.
Ch.
Overview Basic Advanced
9.4.1
1 Select Language
x
9.4.2
3 Process Alarm
x
9.4.3
4 Temp Unit
x
9.4.4
5 Buzzer
x
x
9.4.5
7 Basic PIN Code
x
9.4.6
8 Short Dist lim.
x
9.4.7
9 Long Dist lim.
x
9.4.8
15 Temperature Ctrl
x
9.4.9
16 Disp. Contrast
x
x
9.4.10 18 Distance
x
9.4.11 23 Filter Time
x
9.4.12 25 Contin. Mode
x
9.4.13 36 Advanced PIN Code
x
9.4.14 37 Dist Unit
x
9.4.15 43 Advanced Code
x
9.4.16 44 Reset Prog. settings
x
9.4.17 45 Reset Filter Timer
x
9.4.18 46 Statistics
x
x
9.4.19 47 Perform Self test
x
x
9.4.20 48 Program Time Scale
x
9.4.21 49 Proc. Alarm Temp
x
available in