6
Getting Started
Model 101303-50
Warning: This unit is designed to deliver maximum output
voltage and amperage to the load at 100% controller
output. If the load is designed to run at lower than
the maximum output capability, the controller must
be adjusted to limit the voltage to this load. The
factory has limited the output of this controller as per
the controller settings listed in the Factory
Configuration in Appendix A.
It is likely that the user will need to change these
settings to reach desired operating temperatures and
ramping rates. Consult the Series 96 Manual for
necessary settings.
3.
Plug the unit into a power receptacle.
4.
Do not
connect the system heater at this time.
Warning: This instrument is designed for use with input
voltage, current and frequency as specified in the
Factory Configuration in Appendix A only; serious
equipment damage and/or injury will occur if it is
connected to improper power.
3. Power up the instrument
1.
Turn the instrument on using the front panel switch.
2.
After a few seconds the sensor temperature will be visible in the upper
display of the control module (in Celsius) and the setpoint will be visible in
the lower display. Verify that this sensor temperature appears reasonable.
4. Set over-temperature alarm set point
The over-temperature alarm setpoint of the limit module must be set. For more
information about the user interface, see
Limit Module
Displays and Menu
Navigation above
.