5
Getting Started
Model 101303-50
Getting Started
Getting started using your new HeatWave Labs Temperature Controller should be
quick and easy. The
Error Messages and Troubleshooting
section of this manual
describes errors that may be encountered during this process. For more details about
the operation of the control module, consult the attached
Watlow
®
Series 96 user’s
manual.
This manual is also available for download at the HeatWave labs web site,
www.cathode.com
or at
http://www.cathode.com/pdf/series96man.pdf
.
To ensure the quickest and most reliable startup, please follow the eight steps
below in the order shown.
1. Unpack the instrument
Be sure that you received the following items in good condition. If there was any
damage in shipping, please keep your shipping containers and contact the factory
immediately.
•
Model 101303-50 instrument chassis
•
Power Cord
•
HeatWave Labs p/n 101960-01 (FWH-020A6F) 20A, Fast Acting
Semiconductor Fuse (Installed)
•
Quencharc 102660-03 (not installed)
•
Manuals: HeatWave Labs Model 101303-50,
Watlow
®
Series 96, and
Watlow
®
Series SD user’s manuals
•
HeatWave Labs Catalog
2. Connect the instrument
Warning: This instrument is designed for use with input
voltage, current and frequency as specified in the
Factory Configuration in Appendix A only; serious
equipment damage and/or injury will occur if it is
connected to improper power.
Caution: Air vents are located on top, bottom and rear of the
instrument. Airflow must not be restricted at these
locations.
1.
Make sure the switch on the front panel is in the off position.
2.
Plug your process thermocouple, Type K, into the control sensor receptacle
on the rear panel. (If you need to add connectors to your sensors, do it now.
Remember thermocouples are polarity sensitive, for Type-K the yellow wire
is positive.)