Fan is not working properly
Check whether medication cup for damage.
Check whether the transducer is clean.
Check whether the fan is firmly and properly
mounted on the motor shaft.
Replacement Parts
Replacements parts are available through your local retailer or they can be ordered by mail.
Your Ultrasonic Nebulizer can operate with the use of a rechargeable battery when AC power is not
available. The life of the battery is approximately 2 years. However, battery life will wary depending on
usage, storage and charging conditions
To charge the battery
1.Plug the adapter into a standard AC outlet.
2.Insert the adapter plug into the unit’s AC port, fig.1
Note:
For first time use or after storing the battery for a period of time, charge the battery at least 24 hours
and thereafter 4 to 12 hours depending on the length of usage.
A fully charged battery will last for 20-30 minutes of continuous use.
To prolong the life of the battery, the charging current is designed at the minimum level . Therefore,
it is recommended to charge the battery after each use and let it charge overnight.
To use the battery
1.Insert the battery's
plug into nebulizer's power receptacle.
Fig.2
fig.2
2.Begin the use the Nebulizer as described in the Operating Your Nebulizer section of this guidebook.
3.If the nebulizer's indication lamp changes to red color, it indicates that the battery is in low-voltage
condition and needs to be charged.
4.After use, dispose excess medication and water before disconnecting the battery.
Battery Caution
Do not use the battery with equipment other than your Ultrasonic Nebulizer or attempt to modify it for other
use.
Do not store the battery in direct sunlight high temperature or high humidity.
Avoid placing the battery on metal or wet surfaces or inside a metal box. This may short-
circuit the battery’s
positive and negative contact causing the battery to heat up or explode
Unit specifications
Name or model number
Power
Ultrasonic Nebulizer SHL-US10
220-240VAC 50Hz