Actions to be done by carer
Daily
Weekly
Check operation of side rails
Check patient handset and cable
Check ACP and cable
Visually check castors
Visually check power supply cord and mains plug
Check the mattress for damage and fluid ingress
Examine the lifting pole, belt and handle
Check for correct fitting of the locking pin on removable side
rails and extendable head/foot boards.
Check the Mechanical Quick Release of the backrest
(Mechanical CPR), test it with empty bed (please notice that
at empty bed the gas pump will oppose to the descent of
the backrest)
Actions to be done by qualified personnel.
Yearly
By carer: keep in mind also during daily use the occurrence of signs of malfunction
and/or suspected degradation of the original conditions of the device, reported by
these tests, so that a preventive intervention would be possible.
Carry out a full test of all electrical bed positioning functions (backrest,
height, tilt, etc.)
Check that the linen tray remains in its closed position when maximum
foot down tilt is applied.
Check that the bed operates correctly using the backup battery as described
in the next section
Battery test.
Check operation the of castors, paying special attention to braking and
steering functions
IFU-HBB500-ENG-2021#ed02-May