Operation
ELECTRIC CPR
:
If for emergency reasons is needed to flatten the mattress platform and
all the sections out, then press and hold the CPR button.
The CPR button overrides the function lockout settings.
[For faster emergency procedure directly apply the MECHANICAL CPR
device to drop the backrest.].
BATTERY CHARGE INDICATOR:
If the ACP battery indicators flash ORANGE, the battery is charging.
If the ACP battery indicators are steady ORANGE, the battery is fully
charged. The more battery indicators are steady, the more the battery is
charged (maximum three indicators).
BACKREST SECTION
:
these buttons raise and lower the backrest (upper button to raise, lower
button to lower).
The backrest reaches an angle approximately 50° above the horizontal.
THIGH SECTION
:
These buttons raise and lower the thigh section (upper button to raise,
lower button to lower).
When the thigh section is raised from the flat position (32°), the calf
section will be in the Fowler position (7° angled downwards).
MATTRESS PLATFORM HEIGHT
:
these buttons raise and lower the mattress platform (upper button to
raise, lower button to lower).
Minimum height: 43cm
Maximum height: 79cm
AUTO-CONTOUR:
The Auto-Contour upper button simultaneously raises the backrest and
thigh sections to provide upright patient profiling; the raised thigh section
prevents the patient sliding down the bed.
The Auto-Contour lower button returns the mattress platform to a flat
position.
REVERSE TRENDELENBURG:
This button lowers the foot end of the mattress platform (maximum 12°).
TRENDELENBURG:
This button lowers the head end of the mattress platform (maximum
12°).
AUTO-CHAIR (cardiac-chair)
:
The Auto-Chair button simultaneously raises the backrest (50°) and thigh
sections (22°), then lower the foot end of the mattress platform into a
chair position.
IFU-HBB500-ENG-2021#ed02-May
42
Summary of Contents for HBB500
Page 6: ...General Warnings Continued on next page IFU HBB500 ENG 2021 ed02 May 6 ...
Page 7: ...Continued on next page IFU HBB500 ENG 2021 ed02 May 7 ...
Page 8: ...IFU HBB500 ENG 2021 ed02 May 8 ...
Page 47: ...Duty Cycle Lockout Duty cycle is 10 MAX 2min ON 18min OFF IFU HBB500 ENG 2021 ed02 May 47 ...