Figure 13
4. Insert the Bi-Polar fixture into the end of the black adapter.
5. Insert the Bi-Polar instrument into the top of the Bi-Polar fixture.
(Fig. 14)
.
Figure 14
6. Turn on the insulation tester unit on and set voltage to 2.8- ± 0.3 kV.
7. Touch the bare tip of the Bi-Polar forceps to ensure that the alarm sounds, and
the system is operational.
8. Using the brush, slowly brush from top to bottom of each
(Fig. 15)
:
a. On the outside of left side of tine
b. On the inside of left side of tine
c. On the inside of the right tine
d. On the outside of the right tine.
Figure 15
9. If the alarm sounds, the Bi-Polar has a fault. Follow standard facility procedures
for a defective electrosurgical instrument. (Note: place a repair tag on all
instruments that fail testing.)
Interpretation of Test Results
N/A
Contraindications of Test Results
N/A
Documentation
N/A
Special Warnings and Cautions
●
This should be used in the Sterile Processing area
ONLY
during assembly and
inspection.
●
The Lithium Polymer (LiPo) Battery can only be replaced at the McGan facility.
DO NOT
attempt to replace the battery.
●
(Only) use the charger that comes with the unit.
Document Version
4