Figure 6
2. If using the MM513 Kit, attach the HV Red wire to the pin on the side of the
Saddle Block or connect red port to the top of the unit directly to the side pin.
Make sure controls face up.
(Fig. 7)
.
Figure 7
3. Take the clamp on the green ground wire and attach it to the conductive core of
the instrument under test.
(Fig. 8)
.
Figure 8
4. Turn the base unit on and set the voltage to 2.8- +/- 0.3 kV.
(Fig. 9)
.
Figure 9
5. Push the ESI under test through the LSE ring electrode slowly (approximately
three [3] feet every four [4] seconds).
6. Alarm will sound when the ESI (the bare tip of the instrument) is first inserted
into the electrode.
7. Alarm sounds/LED flashes if fault is found in the coating indicating a fault with
the instrument.
Using the Tri-Hole Electrode:
1. Same setup as shown above, except turn the voltage to 4.2- ± 0.3 kV.