Healthmark MM513-100 Instructions For Use Manual Download Page 5

4. Use 4.2 ± 0.3 kV when using the Tri-Hole electrode.

Figure 4

5. Push the ESI under test through the LSE ring electrode slowly (approximately 

three [3] feet every four [4] seconds).

6. The alarm will sound when the ESI (the bare tip of the instrument) is first 

inserted into the electrode.

7. After the test is completed:

a. Turn the base unit off and remove the clamp end from the unit under 

test.

b. Remove the electrode from the probe wire and remove the ground wire 

and probe wire from the base unit.

c. Properly store the unit and accessories away.

8. Follow the hospital’s procedure policy with regards to the instrument under test.

(Note: The unit should always be switched off prior to removing or repositioning of the 
ground lead, the HV red wire, or the saddle block. If the unit is on and you touch the 
ground lead (clamp end) and the probe end of the base unit at the same time, you will 
receive a very mild “tingle”.  To remove the possibility of receiving the “tingle”, always 
use surgical gloves when handling the leads. You can hold the saddle block from the top 
or the sides—just don’t touch the connection points. 

[Fig. 5]

).

Figure 5

For Round ESI, such as laparoscopic:

1. Insert the chosen electrode into the proper slot in the top of the Saddle Block. 

Make sure the pin is securely placed in the hole. 

(Fig. 6)

.

Document Version

4

Summary of Contents for MM513-100

Page 1: ...rrent output 0 1 mA 0 0001 A at probe Short circuit test current 0 1 mA max Power supply 1800 mAh Li Polymer Battery Probes o Medical style brass wire eight 8 mm wide brush trim length of two 2 mm NOT...

Page 2: ...Insulation Tester MM513 100 Bi Polar Fixture Black Connector MMBPT 190 Ring and Brush Electrodes MMLSE 0029 Ground Wire with Alligator Clip Tri Hole Electrode MMTRI 0022A Saddle Block MMSBT 170 Docum...

Page 3: ...ons N A Packaging Conditions N A Shelf Life N A Instructions for Using Product Description of Use s For testing insulation integrity of electrosurgical instruments Preparation for Use 1 Turn the unit...

Page 4: ...f the unit under test See Operational Guide for Saddle block Diagrams drawings pictures N A Steps for Use of Product 1 Remove the insulation tester unit and accessories from the carrying case 2 Take t...

Page 5: ...pital s procedure policy with regards to the instrument under test Note The unit should always be switched off prior to removing or repositioning of the ground lead the HV red wire or the saddle block...

Page 6: ...st Fig 8 Figure 8 4 Turn the base unit on and set the voltage to 2 8 0 3 kV Fig 9 Figure 9 5 Push the ESI under test through the LSE ring electrode slowly approximately three 3 feet every four 4 secon...

Page 7: ...sert the end of one tine of the Bi Polar forceps into the middle of the brush Fig 11 Figure 11 5 Turn the base unit on and set the voltage to 2 8 0 3 kV 6 Slowly push the Bi Polar forceps away from yo...

Page 8: ...side of tine c On the inside of the right tine d On the outside of the right tine Figure 15 9 If the alarm sounds the Bi Polar has a fault Follow standard facility procedures for a defective electrosu...

Page 9: ...e LED will not illuminate o If the MM513 fails to operate due to battery failure contact Customer Service Always keep the working end of any of the probe electrodes away from your body Do not touch pr...

Page 10: ...rly degradation Avoid contact with such materials It is recommended to calibrate the MM513 base unit P N MM513 110 at least once per year to ensure it is operating at the appropriate voltage Healthmar...

Page 11: ...Reference Documents N A Customer Service Contact Healthmark Industries Company Inc 18600 Malyn Blvd Fraser MI 48026 1 586 774 7600 healthmark hmark com hmark com Document Version 4...

Reviews: