FLEXIBLE INSPECTION SCOPE

healthmark

USER MANUAL & SOFTWARE INSTALLATION GUIDE

Summary of Contents for FIS-005

Page 1: ...FLEXIBLE INSPECTION SCOPE healthmark USER MANUAL SOFTWARE INSTALLATION GUIDE...

Page 2: ...Page 1 of 24 4 24 2018 Flexible Inspection Scope FIS 005SK FIS 005 Product Manual...

Page 3: ...tion Scope 8 Flexible Inspection Scope Viewer Software 8 Selecting the Video Device or Camera 9 Capturing Still Pictures 10 Capturing Video Images 11 Displaying Reference Image 12 Setting File Name 13...

Page 4: ...deo Devices 22 Warranty 23 Specification 24 Contact Information Healthmark Industries Company Inc 33671 Doreka Fraser MI 48026 Toll Free 800 521 6224 Phone 586 774 7600 FAX 586 491 2113 website www hm...

Page 5: ...emitted light or directing it toward others 6 Carefully inspect the external surfaces of the Flexible Inspection Scope and any accessories to assure they are smooth and free of any protrusions or sha...

Page 6: ...n for documentation in hard to see crevices channels and lumens in items areas as small as 2 mm and not visible to the unaided eye The Flexible Inspection Scope can be used with a PC running MS Window...

Page 7: ...nserted into a device during visual inspection 3 Brightness Increase Button Press button to increase illumination intensity 4 Brightness Decrease Button Press button to decrease illumination intensity...

Page 8: ...ckaged before shipping Once the unit is removed from its container it should be carefully inspected for shipping damage If there is any damage contact the shipping carrier and Healthmark Industries im...

Page 9: ...h and free of any protrusions or sharp edges Routine Maintenance There are no user serviceable parts within the Flexible Inspection Scope No user maintenance beyond cleaning is required Refer all serv...

Page 10: ...ill image being displayed in the Main Image Window 4 Open Button Opens the file location File Loc where images are being saved and allows you to select image that is displayed in Reference Image Windo...

Page 11: ...the instructions below for capturing still pictures from the Main Image Window 1 Select the Picture Button on the software FIGURE X 2 Use any of the following options to capture an image a Click Captu...

Page 12: ...elect the Video Button on the software FIGURE X 2 Use any of the following options to start and stop the video a Click the Capture Button on the Flexible Inspection Scope Viewer Software FIGURE X abov...

Page 13: ...the bottom of the software window FIGURE X 7 To stop recording use any method as described in Step 2 above for starting the video Note The Image Capture Button will now read Stop Video FIGURE X while...

Page 14: ...lick the Open Button to display the image in the Reference Image Window Setting File Name Following the steps below allows you to create a file name that will appear after the underscore when capturin...

Page 15: ...RE X in the Flexible Scope Viewer displays image files that are being stored in the File Loc FIGURE X To delete files go the location shown in the File Loc on your computer outside of the Flexible Sco...

Page 16: ...ile inserting into the device If an obstruction hinders the path of the Flexible Inspection Scope gently attempt to manipulate or rotate the device to avoid the obstacle FIGURE X FIGURE X Inserting th...

Page 17: ...ppears white ILLUMINATION 1 The illumination capability of the Flexible Inspection Scope can be verified prior to use Once the scope is connected to your computer with the power on the distal tip of t...

Page 18: ...to follow your institution s specific cleaning and disinfecting or sterilizing procedures in consultation with this manual The user is responsible to ensure that their procedures result in disinfectio...

Page 19: ...iWipes or similar products 3 Secure the Soaking Cap on the Flexible Inspection Scope Camera Connector 4 Place the Flexible Inspection Scope in a bath of enzymatic detergent that has been prepared acco...

Page 20: ...FOR CLEANING section of this manual before proceeding with sterilizing The Flexible Inspection Scope is compatible with STERIS SYSTEM 1E by STERIS Corporation Follow the instructions provided with th...

Page 21: ...nd or internal cables Replace Flexible Inspection Scope Overly bright image White out type reflection Light Intensity is too bright Press Decrease Brightness Button to decrease light output Blurry Ima...

Page 22: ...ble for use with the Flexible Inspection Scope Viewer In the example below the Integrated Webcam is selected 4 Click Add Camera to Remote Capture Exclusions List to disable the selected device 5 Click...

Page 23: ...manufacturer s specifications within this time at no cost to the purchaser It shall be the purchaser s responsibility to return the device directly to Healthmark Industries after receiving a Returned...

Page 24: ...R ANY BREACH OF THE WARRANTY GIVEN ABOVE Specifications Power Source Electrical Power for the Flexible Inspection Scope is supplied by a USB port on a computer Illumination Source LED within the Flexi...

Page 25: ...C to 60 C Pressure 600 hPA to 900 hPA Normal Operation Humidity 0 100 condensing Temperature 5 C to 40 C Waterproof Rating IPX7 Manufacturer Healthmark Industries 33671 Doreka Fraser MI 48045 www hmar...

Page 26: ...cope 45cm 17 7 Weight 18 2 grams 0 64oz Optical Field of View 120 in air Angle of View 0 Distal Camera 1 9 mm 0 075 Distal portion of the Inspection Scope that contains the camera lens Flexible Workin...

Page 27: ...ing the Flexible Inspection Scope install the Healthmark software from the USB Flash drive on a computer System requirements MS Windows versions 7 8 or 10 1 Remove the USB stick and plug into USB port...

Page 28: ...pe Viewer Software screen comes up and click Next 6 Welcome to the Healthmark Scope Viewer Software Setup Wizard screen appears and click Next 7 The License Agreement page comes up and click I Agree i...

Page 29: ...8 Then click Next 9 The Confirm Installation page pops up click Next 10 Installing Healthmark Scope Viewer Software is being installed...

Page 30: ...11 Installation Complete Screen comes up and click Close 12 Once the installation is complete remove the USB...

Page 31: ...nto the top of the Flexible Inspection Scope 14 Plug in the USB Connector to the USB port in your computer VERIFYING OPERATION IMAGE 1 Open the Healthmark Viewer Software on the computer An image from...

Page 32: ...nation capability of the Flexible Inspection Scope can be verified prior to use Once the scope is connected to the computer with the power on the distal tip of the Flexible Inspection Scope will cast...

Page 33: ...6 File Name Box Text box creating a File Name 7 Files Box Displays captured images that are routed and stored in the area being shown in File Loc 8 Settings Button Click to select the video camera and...

Page 34: ...play video devices or cameras that are being detected by your computer 2 Select a device for capturing images using the Healthmark Scope Viewer A The example below shows an Integrated Webcam and USB V...

Page 35: ...n the software Fig 3 Figure 3 2 Use any of the following options to capture an image A Click Capture on the Healthmark Viewer Software Fig 3 B Press the Handpiece Capture Image Button on the Flexible...

Page 36: ...t and stop the video A Click the Capture Button on the Healthmark Viewer Software Fig 5 B Press the Handpiece Capture Button on the Flexible Inspection Scope C Press the spacebar on the computer keybo...

Page 37: ...ved in a file folder titled Reference Images in the designated File Loc Fig 7 Figure 7 2 To display a saved image in the Reference Image Window from you File Loc A Click the Open Button Fig 8 B Click...

Page 38: ...Setting File Name 1 Click Set File Name Button Fig 9 2 Enter the characters that you would like to be included in file name Fig 9A 3 Click the OK Button to set as default name Figure 9...

Page 39: ...save captured images 3 Click OK to set the File Loc for saved files Figure 10 Deleting Saved Files The Files Window Fig 11 in the Healthmark Viewer displays image files that are being stored in the Fi...

Page 40: ...Fig 12 2 Click the Make New Folder Button 3 Create a name for the new folder 4 Click the OK Button to create the new file in File Loc Figure 12 Diagrams drawings pictures Figure 1 Figure 2 Steps for U...

Page 41: ...lexible Inspection Scope is NOT STERILE Follow the instruction for cleaning disinfecting and sterilizing provided in this document Do not autoclave or dry heat the Flexible Inspection Scope Avoid look...

Page 42: ...olution to remove any residues from areas that cannot be reached with the wipes 5 Rinse with low linting wipe moistened with AAMI Utility Water for 45 to 60 seconds dry with low linting wipe Cleaning...

Page 43: ...use Related Healthmark Products Video Inspection Scope with Display optical inspection products Other Product Support Documents ProSys Brochure ProSys Price List Reference Documents Customer Service C...

Reviews: