SPECIFICATIONS  

General 

This digital scale uses highly sophisticated microprocessor technology. Each precision scale is 
designed to provide accurate, reliable and repeatable weight measurements.  This scale uses 
motion-sensing weighing technology to determine the weight of a moving patient.   

The 3001KL-AM / 3001KL-AMX displays weight in pounds (lb) or kilograms (kg). Refer to page 
9 to learn more about configuring the weight measuring unit setting. The 3001KG-AM / 3001KG-
AMX displays weight in kilograms only.

  

The scale can be powered using an AC adapter (ADPT30 included), or by 6 D batteries (not 
included). 

 

Antimicrobial Powder Coating 

Unique to this scale, the handrails and transport handle have been fabricated with a permanent, 
antimicrobial powder coating.  The antimicrobial coating utilizes silver ionic technology to inhibit 
the growth of microbes.   

Scale Specifications 

Capacity and 
Resolution 

3001KL-AM / 3001KL-AMX: 1000 lb x 0.2 lb  (454 kg x 0.1 kg) 

3001KG-AM / 3001KG-AMX: 454 kg x 0.1 kg 

Power Requirements 

100-240V AC, 50-60Hz to 9V DC Adapter (Type: UE15WCP 9 VDC @ 500 

mA), Item # ADPT30 (included) or 6 D batteries (not included) 

Environmental 

Operating temperatures: 50°F to 95°F (10°C to 35°C) 
Storage temperatures: 30°F to 125°F (0°C to 50°C) 
Maximum Humidity: 85% RH 

Platform 

Dimensions 

Le

ngth: 15 3/4” (460 mm) 

Width: 22” (560 mm) 
Height: 2 5/8” (60 mm) 

Assembled 
Dimensions 

3001KL-AM / 3001KG-AM                    3001KL-AMX / 3001KG-AMX:  

Depth: 33 

1/8” / 843 mm                        Depth: 34 1/2” / 841 mm 

Width: 22

” / 559 mm                               Width: 22” / 559 mm 

Height: 52” / 1321 mm                            Height:52” / 1321 mm 

 
 
Scale information can 

be found by pressing           and selecting “Scale Info” in the submenu.  This 

screen provides model number, capacity, calibration weight, the most recent field calibration date, 
number of field calibrations and software version. 

 
 

DEFINITION OF SYMBOLS 

 

 

 

 

 
Scale surface is applied part, Type B.   

Help 

Attention, Consult Accompanying Documents 

Type B Applied Parts 

Type: UE15WCP 
9 VDC @ 500 mA 

Caution 

Summary of Contents for 3001KG-AM

Page 1: ...MX Antimicrobial Digital Platform Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM3001KL AM_3001KL AMX_3001KG AM_3001KG AMX_E_R...

Page 2: ...al 5 Assembly Instructions 6 Set up Instructions 9 Everlock Instructions 13 Operating Instructions 14 Maintenance and Cleaning 20 Calibration 21 Troubleshooting 23 Warranty 24 Note This scale has been...

Page 3: ...cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ensure that th...

Page 4: ...ale Specifications Capacity and Resolution 3001KL AM 3001KL AMX 1000 lb x 0 2 lb 454 kg x 0 1 kg 3001KG AM 3001KG AMX 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15W...

Page 5: ...se contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal This Health o meter Professional scale must be disposed of prope...

Page 6: ...king materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts...

Page 7: ...e completely hand tightened 2 Turn the platform over so it is facing up Locate the handrail and untie the load cell cable connector Position the handrail over the handrail supports on the platform and...

Page 8: ...d with the four mounting holes on the display support Using the 4 included M4 x 0 7 x 10mm screws fasten the display head onto the display support using a Phillips head screwdriver to tighten the scre...

Page 9: ...ead by unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professi...

Page 10: ...the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button to briefly disp...

Page 11: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Page 12: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Page 13: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Page 14: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Page 15: ...by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 4 To rewei...

Page 16: ...may take several seconds for the scale to lock onto the weight 3 After obtaining the patient s weight and with the patient remaining on the scale press 4 The scale will prompt entry of the patient s...

Page 17: ...he tared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as...

Page 18: ...nt The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain...

Page 19: ...parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Page 20: ...and Disinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit...

Page 21: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Page 22: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Page 23: ...n the feet on the underside of the platform to ensure all 4 feet are properly balanced 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing proce...

Page 24: ...ty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make...

Reviews: