OPERATING INSTRUCTIONS (CONT)

 

Parent/Child Function 

The Parent/Child function on Health o meter

®

 Professional scales provides the convenience of 

weighing a small child or infant while being held by a parent. Follow these simple steps to 
subtract the parent’s weight from the total so only the child’s weight is displayed.  
 

1. 

Wait until “0.0” appears on the display. 

 

 

2.  Ask the parent to step onto the scale without the child. The 

display will show a value for t

he parent’s weight.  

 
 
 

3.  With the parent remaining on the scale press          . The scale 

will zero, and the word “TARE” will appear at the bottom 
center of the display along with the parent’s tared weight.  

4. 

Carefully place the child into the parent’s arms. The scale will 
automatically deduct the weight of the parent and only display 
the child’s weight.  

 

 
Note:

 

For an alternate method, after taring the parent’s weight 

in step 3, have the parent step off of the scale. With no weight 
on the scale, the display will show the tare value as a negative 
amount. Have the parent to step back onto the scale while 
holding the child. The scale will automatically deduct the 
weight of the parent and only display the child’s weight.  

 

5.  The tared value will remain on the display until changed, 

cleared, or until the scale is powered off. To remove the Tare, 

press          .. 

The word “TARE” will disappear from the bottom 

of the display, and the tare value will be cleared from the 

display. The scale will resume normal operation. 

 
 
 
 
           
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

Tare 

Tare 

Summary of Contents for 3001KG-AM

Page 1: ...MX Antimicrobial Digital Platform Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM3001KL AM_3001KL AMX_3001KG AM_3001KG AMX_E_R...

Page 2: ...al 5 Assembly Instructions 6 Set up Instructions 9 Everlock Instructions 13 Operating Instructions 14 Maintenance and Cleaning 20 Calibration 21 Troubleshooting 23 Warranty 24 Note This scale has been...

Page 3: ...cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ensure that th...

Page 4: ...ale Specifications Capacity and Resolution 3001KL AM 3001KL AMX 1000 lb x 0 2 lb 454 kg x 0 1 kg 3001KG AM 3001KG AMX 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15W...

Page 5: ...se contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal This Health o meter Professional scale must be disposed of prope...

Page 6: ...king materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts...

Page 7: ...e completely hand tightened 2 Turn the platform over so it is facing up Locate the handrail and untie the load cell cable connector Position the handrail over the handrail supports on the platform and...

Page 8: ...d with the four mounting holes on the display support Using the 4 included M4 x 0 7 x 10mm screws fasten the display head onto the display support using a Phillips head screwdriver to tighten the scre...

Page 9: ...ead by unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professi...

Page 10: ...the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button to briefly disp...

Page 11: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Page 12: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Page 13: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Page 14: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Page 15: ...by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 4 To rewei...

Page 16: ...may take several seconds for the scale to lock onto the weight 3 After obtaining the patient s weight and with the patient remaining on the scale press 4 The scale will prompt entry of the patient s...

Page 17: ...he tared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as...

Page 18: ...nt The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain...

Page 19: ...parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Page 20: ...and Disinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit...

Page 21: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Page 22: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Page 23: ...n the feet on the underside of the platform to ensure all 4 feet are properly balanced 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing proce...

Page 24: ...ty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make...