Health O Meter 3001KG-AM User Instructions Download Page 14

14 

 

OPERATING INSTRUCTIONS 

Key definitions can also be found by pressing            

and selecting “Key Definitions” in the 

submenu.   
 
 

 

 
 
 
 
 
 
 
 
 

 

Keypad

 

Function 

Description 

 

Powers scale on and off. 

 
 
 

Zeroes the scale prior to a weighing. Also used as a screen navigation key to exit a 
display menu. 

 

Holds the weight value on the display until the button is pressed again to clear the 
value.   

 
 
 

Prompts entry of height data to calculate patient’s Body Mass Index (BMI). 

 
 
 

Recalls the previous weight. 

 

TARE-Deducts the weight of an object on the platform.  
PRETARE-When held for 3 seconds, prompts entry of a weight value to be 
deducted from the weight on the platform. Also releases tare weight (returns display 
to zero). 

 
 
 

Toggles between pounds and kilograms.  If weight measuring unit is locked and 
Unit View setting is enabled, press and hold Unit button to briefly display the 
alternate weight measuring unit. 

Note:

 The Unit button is disabled on KG models or 

on KL models with Everlock™ engaged or with the “Unit View” setting disabled.    
 

 
 
 

Allows repeated weighing of the patient without stepping off the scale. 

 

Used to accept height value in BMI mode, and tare value in PRETARE mode.     
Also used to save selections in Set Options menu. 

 

Display navigation keys. 

 

 

 

 

 

Zero/ 

(Clear) 

Hold/ 

Release 

BMI 

Recall 

Tare 

Unit 

Reweigh 

Enter 

 

 

 

 

Weight Unit 

Time 

Battery Power 

Date 

Weight 

Measurement 

Help 

Summary of Contents for 3001KG-AM

Page 1: ...MX Antimicrobial Digital Platform Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM3001KL AM_3001KL AMX_3001KG AM_3001KG AMX_E_R...

Page 2: ...al 5 Assembly Instructions 6 Set up Instructions 9 Everlock Instructions 13 Operating Instructions 14 Maintenance and Cleaning 20 Calibration 21 Troubleshooting 23 Warranty 24 Note This scale has been...

Page 3: ...cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ensure that th...

Page 4: ...ale Specifications Capacity and Resolution 3001KL AM 3001KL AMX 1000 lb x 0 2 lb 454 kg x 0 1 kg 3001KG AM 3001KG AMX 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15W...

Page 5: ...se contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal This Health o meter Professional scale must be disposed of prope...

Page 6: ...king materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any sharp object to open the protective inner packaging Parts...

Page 7: ...e completely hand tightened 2 Turn the platform over so it is facing up Locate the handrail and untie the load cell cable connector Position the handrail over the handrail supports on the platform and...

Page 8: ...d with the four mounting holes on the display support Using the 4 included M4 x 0 7 x 10mm screws fasten the display head onto the display support using a Phillips head screwdriver to tighten the scre...

Page 9: ...ead by unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professi...

Page 10: ...the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button to briefly disp...

Page 11: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Page 12: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Page 13: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Page 14: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Page 15: ...by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release press 4 To rewei...

Page 16: ...may take several seconds for the scale to lock onto the weight 3 After obtaining the patient s weight and with the patient remaining on the scale press 4 The scale will prompt entry of the patient s...

Page 17: ...he tared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as...

Page 18: ...nt The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain...

Page 19: ...parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Page 20: ...and Disinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit...

Page 21: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Page 22: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Page 23: ...n the feet on the underside of the platform to ensure all 4 feet are properly balanced 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing proce...

Page 24: ...ty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty Service make...

Reviews: