![Health O Meter 2210KG-AM User Instructions Download Page 11](http://html1.mh-extra.com/html/health-o-meter/2210kg-am/2210kg-am_user-instructions_2113114011.webp)
P/N 82-00013 UM2210KL-AM 2210KL4-AM 2210KG-AM 2210KG4-AM Rev20210805
11
SET UP INSTRUCTIONS (CONT.)
Everlock
®
The Everlock
®
feature allows the measuring unit to be permanently locked on the scale, disabling the
UNIT
button. The Everlock
®
default setting is “disabled”. Follow this procedure to engage Everlock
®
.
Warning: Everlock is permanent and cannot be reversed.
Notes:
Press
CLEAR/REWEIGH
at any time to exit the Everlock
®
procedure.
Once Everlock
®
is engaged the Unt and UdP options will no longer be displayed when scrolling
through the Options menu.
There is no method to undo Everlock
®
without sending the scale back to the manufacturer.
1. While the scale is turned off, press and hold
ENTER/SELECT
then press and release the
button. Hold
ENTER/SELECT
until “AOb” appears on the display. Use the ▲ or ▼ buttons to
advance the display screen until “ELC” appears. Press
ENTER/SELECT
to change the setting.
ELC =d
Default setting with Everlock
®
disabled.
ELC = E Engages Everlock
®
. Once this option is selected, the display will prompt you to begin
the locking procedure. If this is desired, follow the instructions below.
2. After Everlock
®
is selected, “PASS1” will appear on the display. This prompts the user to enter a
password. Press the following keys in this sequence:
HOLD, BEFORE FEED, AFTER FEED, AMOUNT FED
3.
If entered correctly, “ELC=L” (if scale was in LB mode), “ELC=├” (if scale was in KG mode) or
“ELC=g (if scale was in GM mode) will appear on the display.
4.
Use the ▼ and ▲ keys to toggle between (L) Pounds, (├) Kilograms or (g) Grams. When the
desired unit is displayed, press
ENTER/SELECT
and “PASS2” will appear on the display.
5. This will prompt the user to enter the second password. Press the following keys in this
sequence:
AMOUNT FED, AFTER FEED, BEFORE FEED, HOLD
6.
If entered correctly, the scale will display “ELC = L”, “ELC = ├” or “ELC=9” depending on the unit
of measurement that was selected. Everlock
®
is now engaged.