Specifications
20
A1063-6.0
HBM: public
Z6/M
Z6FC3
Z6FD1
Type
Safe load limit (E
L
)
% of
E
max
150
Breaking load (E
d
)
300
Maximum capacity
kg
50
100
200
500
Permissible dynamic load
(vibration bandwidth to DIN 50100)
% of
E
max
100
100
100
100
Nominal (rated) displacement, approx.
mm
0.27
0.31
0.39
0.6
Weight, approx.
kg
0.5
0.5
0.5
0.5
Degree of protection according to
EN60529 (IEC529)
IP 68
(severe test conditions: 1 m wa
ter column;100 h)
Material
Measuring body
Bellows
Screwed cable gland
Cable sheath
Stainless steel
Stainless steel
Stainless steel / Viton
®
PVC
2)
The values for linearity deviation, relative reversibility error and temperature effect on sensitivity
are typical values. The sum of these values is within the cumulative error limits laid down by OIML
R60.