Mounting
Z6/M
A1063-6.0
HBM: public
11
aligned as near to the perpendicular as possible. This
is the case when the foundation and tank joint are
horizontal and the fastening holes on the foundation
and container joint are suitably aligned.
S
The base plate and cover plate should be firmly fas
tened to the foundation and tank respectively.
S
The fitted EEK4 grounding cable is used to protect
against welding currents that can damage the elec
tronic parts of the transducers.
Notice
If the load cell is to be replaced, the upper module plate
can be lifted by max. 1.5 mm.
Lock the nut in place to hold the upper module plate in
this position. Dismounting the tank itself is not required.
The following tightening moments prescribed for the fas
tening bolts must be adhered to when replacing the load
cell:
up to 200 kg max. Capacity
Galvanized version:
34 N·m
Stainless steel version:
23 N·m
500 kg max. Capacity
Galvanized version:
76 N·m
Stainless steel version:
45 N·m
The setting of the overload stop must be checked after
the replacement load cell has been installed. Please refer
to the specifications for the weighing module: ”Setting
range for the overload stop”.
Summary of Contents for Z6/M
Page 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais Z6 M...
Page 3: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais Z6 M...
Page 23: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais Z6 M...
Page 43: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais Z6 M...
Page 63: ...Caract ristiques techniques Z6 M A1063 6 0 HBM public 21...