43
Z4A
HBM
A0697-7.0 de/en/fr
15
Specifications
Force transducer type
Z4A
Data according to VDI standards 2638
Nominal force
F
nom
kN
20
50
100
200
500
Class
1)
00
0.5
Nominal sensitivity
C
nom
mV/V
2
rel. sensitivity deviation
d
c
%
0.1
rel. tensile/compressive force sensitivity
difference
d
zd
%
0.2
Zero signal tolerance
d
s,o
%
0.5
rel. zero point compensation (zero signal
return)
1)
f
o
%
0.008
Rel. range (0.2F
nom
to F
nom
) at:
1)
unchanged mounting position, typically
b
rg
%
0.02
different mounting positions, typically
b
rv
%
0.03
Rel. range of inversion (0.2F
nom
to
F
nom
)
1)
u
%
0.06
0.15
Linearity deviation
d
lin
%
0.02
0.03
Effect of temperature on sensitivity/10 K
by reference to sensitivity
TK
c
%
0.01
Effect of temperature on zero signal/10 K
by reference to sensitivity
TK
0
%
0.015
Effect of transverse forces (transverse
force 10 % F
nom
)
2)
d
Q
%
0.03
Effect of eccentricity per mm
d
E
%
0.01
0.005
Rel. creep over 15 min
d
crF+E
%
0.02
Input resistance
R
e
Ω
>345
Output resistance
R
a
Ω
356
"
0.3
Isolation resistance
R
is
Ω
>5
⋅
10
9
Reference excitation voltage
U
ref
V
5
Operating range of the excitation voltage
B
U,G
V
0.5 ... 12
Nominal temperature range
B
t,nom
°
C
+10...+40
Operating temperature range
B
t,G
°
C
-30...+85
Storage temperature range
B
t,S
°
C
-50...+85
Reference temperature
t
ref
°
C
+22
Max. operational force
(F
G
)
%
150
Limit force
(F
L
)
%
150
Breaking force
(F
B
)
%
250
Static lateral limit force
(F
Q
)
%
30
Summary of Contents for Z4A
Page 2: ...Deutsch Seite 3 26 English Page 27 50 Français Page 51 73 ...
Page 26: ...26 Z4A HBM A0697 7 0 de en fr ...
Page 50: ...50 Z4A HBM A0697 7 0 de en fr ...
Page 74: ...74 Z4A HBM A0697 7 0 de en fr ...
Page 75: ......