QuantumX / SomatXR and CCP / XCP-on-CAN
CANBus
A4461-2.0
HBM: public
67
Step 3 - Export Measurement configuration to a *.dbc
file using CANape:
S
·Select ECU in Logger Configuration (Tools -> Logger
configuration) and click on “Create file” button.
S
=> *.dbc file is created in your CANape project direc
tory
Step 4 – Configure MX471B using *.dbc file and MX
Assistant:
S
·Open MX Assistant and connect to your MX471B
S
·Select Channel 1 of the MX471B and open CAN Bus
settings
S
·Enter Bit rate of your CAN network
S
·Activate Termination if necessary
S
·Select the *.dbc file created in step 3
S
·If ECU is locked by Seed & Key mechanism you
have to select the *skb file.
S
·If ECU is not locked, leave field blank
S
·Click OK button