Application instructions
CLP
A03616_04_YI0_02
HBM: public
9
3
Application instructions
Piezoelectric force washers of the CLP type series are suitable for measuring
compressive forces. Because they provide highly accurate dynamic and quasi-
static force measurements, they must be handled very carefully. Particular care
must be taken when transporting and installing the devices. Dropping or knock
ing the transducer may cause permanent damage.
The transducers are extremely stiff and have a high natural frequency.
Limits for acceptable mechanical, thermal and electrical stress are listed in the
Specifications section on page 19
. It is essential that these are taken into ac
count when planning the measuring set-up, during installation and, ultimately,
during operation.
Notice
Like all piezoelectric force sensors, the piezoelectric force washers of the CLP
series are only exposed to great force (greater than 10% of the sensor capac
ity) if they are connected to a charge amplifier, or short-circuited. If the sensors
are loaded without this measure being taken, voltage flashover can result,
which could damage the sensor.