![HBM MXFS User Manual Download Page 60](http://html1.mh-extra.com/html/hbm/mxfs/mxfs_user-manual_2111808060.webp)
Operation
60
A05595_02_E00_00
HBM: public
MXFS
Strain with temperature compensation
Using a temperature sensor
When using a temperature channel to compensate for the effect of tempera
ture on the strain measurement, it must be ensured that the changes in tem
perature felt by the two sensors is the same. The selected channel for temper
ature compensation with this method must be configured as a temperature
sensor.
Fig. 4.18
Strain with compensation using a temperature sensor
The gauge factor (k) of FBG strain gauges is given in their documentation.
The Temperature Cross Sensitivity (TCS) corresponds to the effect of temper
ature on the strain sensor, meaning the induced strain to the sensor after
installation due to a change of 1°C on its temperature. It is a value given on the
sensor’s documentation.