Connection
HLCM…
A1140-4.0
HBM: public
13
8
Connection
Load cells can be connected either to Carrier‐frequency‐
amplifiers or DC‐amplifiers.
The load cell connection cable should be routed so that
any condensed water or dampness forming on the cable
can drip off. It must not be led to the load cell. In addition,
it must be ensured that no dampness can penetrate the
open end of the cable.
6-wire
circuit
Wiring code
(white)
(black)
(grey)
Excitation (+)
(green)
(red)
Excitation (-)
Sense (-)
Signal (+)
(blue)
Sense (+)
Signal (-)
Shield, connected with housing
1) 1)
When connecting to amplifiers in 4‐wire technique you must connect the wires blue to green
and black to gray. The following deviations occur with
unshortened
cables (3 m):
•
sensitivity -0.2 %
•
TKC -0.01 % / 10 K
Notice
Electric and magnetic fields often cause interference volt
ages to be coupled into the measurement circuit. There
fore:
S
Only use shielded, low capacity measurement cables
(measurement cables by HBM meet these require
ments).
Summary of Contents for HLCM Series
Page 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais HLCM...
Page 3: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais HLCM...
Page 24: ...Specifications of the load cells HLCB 22 A1140 4 0 HBM public HLCM...
Page 25: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais HLCM...
Page 47: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais HLCM...