Electrical connection
C16i
A05516_01_Y00_01
HBM: public
21
For C16i3 (third generation and predecessors)
Power supply table (when using the HBM interface converter set SC232/422B
with an internal consumption of approx. 70 mA):
Num
ber of
load
cells
Sum
current
con
sump
Maximum cable length
between converter and junction box
Nominal voltage 12 VDC
Nominal voltage 15 VDC
Wire cross-section
0.14mm
2
0.25mm
2
0.5mm
2
0.14mm
2
0.25mm
2
0.5mm
2
1
120 mA
352 m
500 m
500 m
500 m
500 m
500 m
2
170 mA
176 m
314 m
500 m
296 m
500 m
500 m
3
220 mA
117 m
210 m
419 m
197 m
352 m
500 m
4
270 mA
88 m
157 m
314 m
148 m
264 m
500 m
6
370 mA
59 m
105 m
210 m
100 m
176 m
352 m
8
470 mA
44 m
79 m
157 m
74 m
132 m
264 m
10
570 mA
35 m
63 m
126 m
59 m
106 m
211 m
12
670 mA
29 m
52 m
105 m
49 m
88 m
176 m
16
870 mA
22 m
39 m
79 m
37 m
66 m
132 m
1)
Incl. HBM interface converter set SC232/422B
Notice
The plug-in power supply unit , included in the scope of delivery for the HBM
interface converter set SC232/422B supplies 15 VDC/530 mA. So it can be
used for operation with max. 8x C16i... load cells.
5.3
Command set C16i...
Please request the separate documentation of our AD104C electronics, if re
quired. Operating instructions are available in German and English language.
Summary of Contents for C16i series
Page 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais C16i...
Page 3: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais C16i...
Page 28: ...Dimensions 26 A05516_01_Y00_01 HBM public C16i...
Page 29: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais C16i...
Page 56: ...Abmessungen 28 A05516_01_Y00_01 HBM public C16i...
Page 57: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais C16i...
Page 83: ...Dimensions C16i A05516_01_Y00_01 HBM public 27...