4
Technical Information
Controller Functions
Troubleshooting
Refer to
www.nortev.com
website for detailed information on troubleshooting Flexineb problems.
Maintenance / Cleaning
Medication Cup:
Cleaning:
It is very important to wash out the Med Cup after each use to avoid contamination or damage
of the metallic vibrating mesh. Protect the electrical connector from getting wet when cleaning by using
the protective cap provided. To clean out the Med Cup open up the soft blue Cap and place 2-3 drops of
mild dishwashing liquid (
do not use dishwasher detergent
) into the Cup and fill with distilled water.
Close the soft blue Cap and shake the Cup vigorously for about 15 seconds and squeeze the soft Cap to
push water through the mesh whilst shaking. Open the Cap and rinse out from both sides with distilled
water. Shake off excess water and let air dry before re-use.
Disinfection:
To disinfect the outside surfaces of the Med Cup use anti-bacterial wipes.
CAUTION:
Do not immerse the Med Cup in water or cleaning chemical solutions!
Do not nebulise or put chemical cleaning solutions inside the Cup as they may permanently
damage the metallic mesh or reduce its life!
Do not use sharp instruments to clean the metallic mesh in the Med Cup!
For continuity of treatment it is highly recommended to carry spare Med Cups!
Mask:
Cleaning:
The Mask can be
washed using clean warm tap water with some dish washing up liquid.
Rinse off with warm clean tap water and dry off before re-use.
Disinfection:
First disassemble all parts of the Mask and wash them thoroughly with clean warm tap
water and dish washing up liquid. Rinse the parts off and let them air dry on a clean dry surface.
Disinfect each part by wiping down thoroughly with anti-bacterial wipes
before re-assembly.
Alternatively the Mask parts can be soaked in Nortev approved disinfectant chemicals.
Refer to separate information sheet or
www.nortev.com
for details on recommended disinfectants.
Check
www.nortev.com
for updates and video instructions on Cleaning & Disinfecting Flexineb.
Controller:
To maintain the function of the electronic controller it must be used and stored in dry conditions.
To maintain the battery function it must be charged when the Operation LED turns Red.
CAUTION:
Do not immerse the controller in water or any liquid as this will permanently damage the electronics!
Warranty
Mask Assembly (excluding valves), Controller (excluding battery) and Cable, Power Supply (charger): 24
months. Battery: 6 months. The Medication Cup is a wearing part and its life is very dependent on drugs
nebulised and how well it is maintained therefore it is not covered by warranty.
Damage caused by improper handling or misuse is not covered by warranty.
Compliance
Flexineb complies with the essential requirements of all relevant European Directives and Harmonised
Standards
Customer Service
+353 91 79 97 78 or email
or contact your local distributor
To order replacement parts contact your local distributor or visit our online shop at
www.nortev.com
Postal Address:
Nortev Limited
,
Unit 18
,
Corporate Park
,
Claregalway
,
Co. Galway, Ireland
Check the connections or replace the broken
cable or medication cup to fix the problem.