![Hawk Intelligent 2000 Series Instruction Manual Download Page 21](http://html1.mh-extra.com/html/hawk/intelligent-2000-series/intelligent-2000-series_instruction-manual_2109377021.webp)
21
Overview
10. Specifications
Manufacturer
Klay Insturments
Instrument
Series 2000, Series 2000-SAN,
Series CER-2000 and 2000-Hydrobar
Output
4-20 mA
Optional: HART Protocol
Power Supply
12 - 36 Vdc
Ex-version: 12 - 26,5 Vdc
Accuracy
0,1% of adjusted span
Ranges
Code
Adjustable Span Ranges
Max. Overpressure.
Series 2000
1
0-0,1 bar
0-0,4 bar
6,4 bar
2
0-0,3 bar
0-1,2 bar
10,5 bar
3
0-1 bar
0-10 bar
30 bar
4
0-5 bar
0-30 bar
100 bar
5
0-20 bar
0-100 bar
200 bar
Series 2000 (SAN)/2000-Hydrobar
1
0-0,04 bar
0-0,4 bar
6,4 bar
2
0-0,1 bar
0-1,2 bar
10,5 bar
3
0-1 bar
0-10 bar
30 bar
4
0-5 bar
0-30 bar
100 bar
5
0-20 bar
0-100 bar
200 bar
Series CER 2000
1
0-0,2 bar
0-0,8 bar
5 bar
2
0-0,8 bar
0-2 bar
10 bar
3
0-2 bar
0-10 bar
30 bar
4
0-10 bar
0-40 bar
120 bar
5
0-40 bar
0-200 bar
350 bar
6
0-150 bar
0-320 bar
600 bar
Process Temperature
1
Series 2000
-20
o
C to + 80º C (-4
o
F to 176
o
F)
Series 2000 (SAN)
-20
o
C to +100
o
C (-4
o
F to 212
o
F)
Series CER-2000
-20
o
C to +100
o
C (-4
o
F to 212
o
F) (only gases)
Series 2000-Hydrobar
-10
o
C to + 70
o
C (+10
o
F to 160
o
F)
Temperature Effect
0,015% / K
Ambient Temperature
-20
o
C to +70C (-4
o
F to 158
o
F)
For Ex-version see paragraph 4.4
Damping
0,5 s to 25 sec (0,5 sec = Std. Damping)
Protection Grade
IP 66
IP 68 (for 2000-Hydrobar-cable submersed parts)
Material
Housing
AISI 304
“wetted” part
AISI 316 L
1
For higher temperatures use other kind of pressure transmitters. Contact Hawk Measurement Systems for information.
Series 2000 HART