27
Phenomenon
Reason
Corrective actions
The output image
is not clear.
Objective lens is fuzzy
Clean the mucous membranes and other attachments
on the objective lens by water or cotton
Water droplets or color
bar is found in the Field
Stop use; please send it to service department.
Bending is not
The value of bending
angle under than
standard value
Stop use; please send it to service department.
Pull rod bending the
Clearance in bending
wheel
Resistance of bending
wheel is normal
If the device doesn’t work after you do above corrective actions, please hand this device to HAWK
service and the device must conform to the following requirements.
●
It should clean and sterilize the device before packing it.
●
This device is a precision device, please use strong and shock-proof package to avoid the damage in
the conveyance.
●
Please writing the Symptoms of malfunction. If you have any other requirements, please contact
HAWK service.
6
.
Appendix
6.1 Technical Specifications
Monitor 3.5”
Resolution
960
×
240
Storage SD
Card
Inner power supply
3.7V / 3600mAh battery
Continuous working time
≥
4 hours
Size
100×89×28 (mm)
Monitor 8”
Resolution
800
×
600
Storage SD
Card
Inner power supply
4 pieces battery (18650)
Recommended capacity 2200mAh one battery
Continuous use
≥
4 hours
Size 280×210×50(mm)
Videoscope
Image sensor
1/10
″
CMOS