4.3.5 If you need to record the examination or treatment process, you can take pictures or video.
Press the photo button on the monitor, confirm or button 1 on the endoscope can take pictures./
Press video recording button or press the button 2 on video can record video.
4.3.6 After each use, turn off the power switch, unplug the monitor cable and remove the
endoscope.
4.3.7 When the display power is low, use DC5V / 3A power adapter, charging time is approximately
5-7 hours after full charge, indicator turn off. After charging, please disconnect the power adapter.
Or remove the battery alone charge the battery
Warning
●
The operator and assistant should wear rubber gloves.
●
Do not look directly at the light emitted by the device, to avoid hurt the eyes.
●
Do not operate videoscope for a long time. Since the light emitting lead to high
temperatures and light, the patient or operator may hurt, in order to prevent injury, it
should not take long observation time of videoscope tip near the mucosa or mucous
membrane
●
If the device in question is running, the power switch must be turned off immediately,
unplug the cable, interrupt the operation. And readjust the device by the technician, if it
still have problems, please contact HAWK service department for repair.
Note
Do not bend too hard, if the operator feel the resistance is too large, bending is not
in place or cannot bend, the videoscope should be returned for repair.
●
Do not adjust the wheel when insertion part in coil condition, to avoid damage the
internal parts.
●
Do not over force when insert the forceps, if you found large resistance, do not force
please identify the reasons or return to factory for repair.
●
Do not insert forceps when distal part in coil condition, to avoid damage the internal
parts.
●
Do not use non-DC5V power adapter to charge the display, otherwise it will damage
the monitor.
●
Fixed cover for gas sterilization when in long-term storage, and do not bend the
insertion part.