Page 12
8
Technical data
8.1
Characteristics and specification
Reference temperature 23 °C
Display:
Digital (12.7209.000)
3
1
/
2
-digit liquid-crystal display
Analog (12.7215.000)
Moving-coil instrument with mirrored scale,
polarity display by means of light-emitting diodes
3 mm
Measuring ranges:
Range 1
-20 kV/m to +20 kV/m
Range 2
-200 kV/m to +200 kV/m
Range 3
-2000 kV/m to +2000 kV/m
Measuring distance
30 mm
Analog output:
Output voltage
-2 V to +2 V ( Imax = 0.2 mA )
Measuring accuracy:
Digital
±
10 %
Analog
±
15 %
Over-range:
Digital Display
indicates
I.
Analog
Pointer exceeds the upper scale value
Battery check:
Digital Display
indicates
LO
BAT
Analog Both
polarity
indicating diodes light up
8.2
Supply voltage
Mains:
Voltages
115 V / 230 V (+6 % / -10 %)
Frequency range
50 – 60 Hz
Frequency tolerance range
47,5 – 66 Hz
Power input
10 VA
Rechargeable battery operation:
Number of cells
4 NiMH round cells (AA)
Cell voltage
1,2 V
Battery operation time
approx. 5 h at 800 mAh cell capacity
Summary of Contents for Statometer II