Before measuring, check the SCD. Check the
function of the display and the state of the
battery.
Press and hold the start button. The display will
show all segments briefly (Fig. 5).
Fig. 5
The LOBAT indicator will flash to alert you to a
low voltage of the block battery.
If the LOBAT indicator is displayed
continuously, the voltage is too low for accurate
measurement. No reliable measuring value can
be provided. Replace the block battery (see
page 14: “Replacing the block battery“).
NOTE:
The SCD must be switched off so that all
segments of the display can be shown while
holding the Start button. The SCD will switch
off automatically after 15 s.
Visually inspect the SCD for any contamination.
To obtain a reliable measuring value, the spacer
pins and sensor plate must be dry and free of
grease (see page 14: “Cleaning the SCD“).