HATCH X-RAY Operating Manual
61
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 60601-1-2:2014
IEC 61000-4-
11:2004
0% UT for 0,5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270°
and 315°
0% UT for 1
cycle At 0°
70% UT for
25/30 cycle
At 0°
0% UT for
250/300 cycle
0% UT for 0,5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0% UT for 1
cycle At 0°
70% UT for
25/30 cycle
At 0°
0% UT for
250/300 cycle
Mains power
quality should be
that of a typical
commercial
or hospital
environment. If
the user of the
Hatch X-Ray
Variants requires
continued
operation during
power mains
interruptions, it
is recommended
that the Hatch
X-Ray be
powered from an
uninterruptible
power supply.
Power frequency
(50/60 Hz) magnetic
field
IEC 60601-1-2:2014
IEC 61000-4-8:2009
3 A/m 50Hz
OR 60Hz
3 A/m 50Hz
OR 60Hz
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
Table C2
: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity – For all
EQUIPMENT and SYSTEMS
Summary of Contents for X-RAY
Page 1: ...OPERATING MANUAL ...
Page 3: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 37: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 45: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 57: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 59: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 65: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...