HATCH X-RAY Operating Manual
60
Guidance and manufacturer’s declaration – Electromagnetic
immunity – For all EQUIPMENT and SYSTEMS
Immunity test
EN 60601 test
level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
Discharge (ESD)
IEC 60601-1-2:2014
IEC 61000-4-2:2008
±8 kV contact-
Direct
±8 kV contact-
Indirect
± (2, 4, 8, 15)
kV air
±8 kV contact-
Direct
±8 kV contact-
Indirect
± (2, 4, 8, 15)
kV air
Floors should be
wood, concrete
or ceramic
tile. If the floor
is covered
with synthetic
material, the
relative humidity
should be at
least 30%.
Electrical fast
transient / burst
IEC 60601-1-2:2014
IEC61000-4-4:2012
±2 kV for
power supply
lines 100kHz
repetition
frequency
±1 kV for
Signal lines
100kHz
repetition
frequency
±2 kV for
power supply
lines 100kHz
repetition
frequency
±1 kV for
Signal lines
±1 kV for
Signal lines
100kHz
repetition
frequency
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge
IEC 60601-1-2:2014
IEC 61000-4-5:2005
± 1 kV
Differential
mode
± 2 kV
Common
mode
± 1 kV
Differential
mode
± 2 kV
Common mode
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Summary of Contents for X-RAY
Page 1: ...OPERATING MANUAL ...
Page 3: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 37: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 45: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 57: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 59: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...
Page 65: ...THIS PAGE IS LEFT BLANK INTENTIONALLY HATCH X RAY Operating Manual ...