No.:GC-JS-21 Version: 01/00
Effective date
:
2020-03-30
20
:
The user sets whether to print records or not, and the
page after entering is like this
:
On: printer prints records, SD card stores records;
Off: printer does not print records, SD card stores
records;
6.1.4.6
:
The user can change or set the drying time of 4 programs.
:
Adjustment range: 0 ~ 60 minutes.
Drying time of 134
℃
solid program is not
adjustable
If the user finds that the drying effect after
sterilization is not good, the user can increase
the
drying time
of
the
corresponding program here.
The drying time can be changed according to the
time display.
6.1.4.7
:
The user can set whether to enter with a password when starting up, and set the
entry password
:
Press the "Enable" button to set whether this
function is enabled:
Enable: This function is turned on, and you
need to enter the password when you turn on
the computer to enter.
Not enabled: This function is not turned on, and
it is directly entered when starting up.
Press the password display to change the
password.
6.1.4.8
:
Enter here to view the
information of this device
6.1.5
Users cannot enter, only the manufacturer or the manufacturer's representative. Password required.
6.1.6
The user can enter to view and make up the sterilization record for nearly 20 times, see
7.12 for details.
6.1.7 To-be-run page and running interface
6.1.7.1 After the program is selected, enter the interface to be run, the interface is as follows
6.1.4.5